AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Received Jan 19, 2026 · Event occurred Jan 1, 2017
Report 3005099803-2026-00237 · MDR key 24112036
Device
Generic name
Pancreatic Stent, Covered, Metallic, Removable
Manufacturer
Boston Scientific CorporationProduct problems
- Use of Device Problem
Patient
90 YR · Male
- Peritonitis
Narrative
Additional Manufacturer Narrative
BLOCK B3: APPROXIMATED BASED ON THE DATE THE STUDY WAS CONDUCTED. BLOCK D4, H4: THE LITERATURE ARTICLE DID NOT PROVIDE THE SUSPECT UPN AND LOT NUMBER. BECAUSE THE PRODUCT IS UNKNOWN AT THIS TIME, WE ARE UNABLE TO PROVIDE THE COMPLETE UNIQUE IDENTIFIER (UDI) # AND OTHER PRODUCT SPECIFIC INFORMATION. IF ADDITIONAL DETAILS BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. BLOCK G2: LITERATURE SOURCE: MANUEL PEREZ-MIRANDA, ET. AL., ": LONG- TERM CLINICAL SUCCESS OF ENDOSCOPIC ULTRASOUND- GUIDED GASTROENTEROSTOMY IN BENIGN GASTRIC OUTLET OBSTRUCTION" DIGESTIVE ENDOSCOPY, 2025; 37:1198-1206. HTTPS://DOI.ORG/10.1111/DEN.15087. BLOCK H6: IMDRF PATIENT CODE E1024 CAPTURES THE REPORTABLE PATIENT COMPLICATIONS OF SPONTANEOUS BACTERIAL PERITONITIS. IMDRF IMPACT CODE F2202 CAPTURES THE ADDITIONAL ENDOSCOPIC PROCEDURE FOR STENT PLACEMENT. IMDRF IMPACT CODE F2301 CAPTURES THE USED OF ANOTHER METAL STENT THAT WAS PLACED THROUGH THE LAMS.
Description of Event or Problem
BOSTON SCIENTIFIC CORPORATION BECAME AWARE OF AN EVENT INVOLVING AN AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM THROUGH AN ARTICLE TITLED, "LONG- TERM CLINICAL SUCCESS OF ENDOSCOPIC ULTRASOUND- GUIDED GASTROENTEROSTOMY IN BENIGN GASTRIC OUTLET OBSTRUCTION", BY MANUEL PEREZ- MIRANDA, ET AL. THE STUDY AIMED TO EVALUATE LONG-TERM CLINICAL OUTCOMES OF EUS-GE IN BENIGN GASTRIC OUTLET OBSTRUCTION (BGOO). ACCORDING TO THE ARTICLE, A MULTICENTER RETROSPECTIVE STUDY WAS CONDUCTED ACROSS NINE CENTERS INVOLVING PATIENTS WHO UNDERWENT EUS-GE FOR THE TREATMENT OF BGOO. PATIENTS SCHEDULED FOR EUS-GE WITH LAMS BETWEEN JANUARY 2017 AND JUNE 2023 WERE CONSIDERED ELIGIBLE. THERE WERE 62 PATIENTS INCLUDED FOR ANALYSIS. MOST CASED OF BGOO WERE RELATED TO PANCREATIC DISEASE. ACCORDING TO THE ARTICLE, A 90-YEAR-OLD MALE WITH CHRONIC PANCREATITIS INDUCES DUODENAL STRICTURE AND SECONDARY ASCITES DUE TO HYPOALBUMINEMIA DEVELOPED RECURRENT GOO 94 DAYS POST-LAMS DEPLOYMENT. A TUBULAR SELF-EXPANDING METAL STENT WAS PLACED THROUGH THE LAMS. THE PATIENT WAS ADMITTED DUE TO SPONTANEOUS BACTERIAL PERITONITIS. PLEASE SEE THE REFERENCED ARTICLE FOR FULL DETAILS. NOTE: IT WAS REPORTED THAT THE AXIOS STENT WAS PLACED TO TREAT A BENIGN GASTRIC OUTLET OBSTRUCTION. HOWEVER, THE AXIOS STENT ELECTROCAUTERY-ENHANCED DELIVERY SYSTEM INSTRUCTIONS FOR USE STATE THAT "THE AXIOS STENT IS INDICATED FOR USE TO FACILITATE TRANSGASTRIC OR TRANSDUODENAL ENDOSCOPIC DRAINAGE OF: A PANCREATIC PSEUDOCYST OR A WALLED-OFF NECROSIS WITH GREATER THAN OR EQUAL TO 70% FLUID CONTENT; THE GALLBLADDER IN PATIENTS WITH ACUTE CHOLECYSTITIS WHO ARE AT HIGH RISK FOR, OR UNSUITABLE FOR, SURGERY; AND THE BILE DUCT AFTER FAILED ERCP IN PATIENTS WITH BILIARY OBSTRUCTION DUE TO A MALIGNANT STRICTURE." FURTHERMORE, THE INSTRUCTIONS FOR USE STATE "CAUTION: THE AXIOS IS INTENDED FOR IMPLANTATION UP TO 60 DAYS." THE STENT IS NOT INTENDED TO TREAT A BENIGN GASTRIC OUTLET OBSTRUCTION OR TO REMAIN IMPLANTED FOR MORE THAN 60 DAYS.