inforMED
MalfunctionIOQ

TOTALCARE BARIATRIC PLUS BED

Received Jan 19, 2026 · Event occurred Dec 30, 2025

Report 1824206-2026-00130 · MDR key 24109564

Device

Generic name

Bed, Flotation Therapy, Powered

Model number

P1840D

Catalog number

P1840DCA000037

Product problems

  • Positioning Failure

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT TOTALCARE BARIATRIC CAPITAL (PRODUCT CODE P1840DCA000037, SERIAL NUMBER (B)(6)), HEAD OF BED IS UNABLE TO BE RAISED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED

Additional Manufacturer Narrative

THE BAXTER TECHNICIAN FOUND THAT BOTH HEAD VALVES NEEDED TO BE REPLACED. PER THE BAXTER SERVICE MANUAL THE TOTALCARE BED SYSTEM REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM A SEMI-ANNUAL PREVENTATIVE MAINTENANCE. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE HEAD SECTION SLIDE PIVOTS, SELF-LUBRICATING BEARING, AND SLIDER PIVOT EXTRUSIONS. LOOK FOR DAMAGE, AND MAKE SURE THE HEAD SECTION OPERATES CORRECTLY. REPLACE COMPONENTS AS NECESSARY. A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE HEAD VALVES AND BLEED AIR OUT OF LINES TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.