inforMED
MalfunctionDSA

ECG CABLES AND LEADSETS

Received Jan 17, 2026 · Event occurred Oct 7, 2025

Report 9610816-2026-100098 · MDR key 24105426

Device

Generic name

Cable, Transducer And Electrode, Patient, (including Connector)

Catalog number

989803144991

Product problems

  • Incorrect Measurement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT ECG LEAD SET IS NOT PROVIDING THE CORRECT READINGS. THE DEVICE WAS NOT IN USE ON A PATIENT AT THE TIME OF THE EVENT. THERE WAS NO ADVERSE EVENT REPORTED.

Additional Manufacturer Narrative

A PHILIPS REMOTE SERVICE ENGINEER (RSE) SPOKE TO THE CUSTOMER BIOMED WHO STATED THAT DURING A PREVENTIVE MAINTENANCE CHECK THE LEAD SET WAS PROVIDING INACCURATE READINGS. FURTHER INVESTIGATION FOUND THE LEAD SET WAS DAMAGED AND NEEDED TO BE REPLACED. REPLACEMENT PARTS WERE ORDERED AND SHIPPED TO THE CUSTOMER SITE. THE INVESTIGATION CONCLUDES THAT NO FURTHER ACTION IS REQUIRED AT THIS TIME. IF ADDITIONAL INFORMATION IS RECEIVED THE COMPLAINT FILE WILL BE REOPENED. REPORTING INSTITUTION PHONE # (B)(6). REPORTER PHONE # (B)(6).