inforMED
MalfunctionNIG

ELECSYS CA 19-9

Received Jan 16, 2026 · Event occurred Dec 24, 2025

Report 1823260-2026-00194 · MDR key 24094698

Device

Generic name

Carbohydrate Antigen (ca19-9) Test System

Manufacturer

Roche Diagnostics

Catalog number

07027028190

Lot number

84908301

Product problems

  • High Test Results
  • Low Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ANALYZER SERIAL NUMBER IS (B)(6). THE QC RECOVERY DATA PROVIDED WAS WITHIN SPECIFICATION. THE ALARM TRACE SHOWED SEVEN SAMPLE CLOT DETECTION ALARMS. THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS CA 19-9 RESULTS FOR 1 PATIENT SAMPLE ON A COBAS E 801 ANALYTICAL UNIT. THE INITIAL RESULT WAS 46.1 U/ML. THE SAMPLE WAS REPEATED TWICE, AND THE RESULTS WERE BOTH 2.00 U/ML WITH A DATA FLAG. NO QUESTIONABLE RESULTS WERE REPORTED OUTSIDE OF THE LABORATORY. THE REPEAT RESULTS WERE DEEMED CORRECT.

Additional Manufacturer Narrative

BASED ON THE QC DATA, A GENERAL REAGENT ISSUE COULD BE EXCLUDED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.