inforMED
MalfunctionFLE

AMSCO

Received Jan 15, 2026 · Event occurred Nov 24, 2025

Report 3005899764-2026-00005 · MDR key 24089486

Device

Generic name

Sterilizer

Catalog number

SR3201110111

Product problems

  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A STERIS SERVICE TECHNICIAN ARRIVED ONSITE FOLLOWING THE REPORTED EVENT TO INSPECT THE AMSCO 400 STERILIZER AND FOUND THE VIEW PORT ON THE STERILIZER'S STEAM GENERATOR WAS DAMAGED. AS THE VIEW PORT WAS DAMAGED, THIS ALLOWED WATER AND STEAM TO LEAK OUT ONTO THE FLOOR. THE TECHNICIAN REPLACED THE VIEW PORT, TESTED THE FUNCTION AND OPERATION OF THE DEVICE AND CONFIRMED IT TO BE OPERATING TO SPECIFICATIONS. THE UNIT WAS RETURNED TO SERVICE, AND NO ADDITIONAL ISSUES HAVE BEEN REPORTED.

Description of Event or Problem

THE USER FACILITY REPORTED THAT STEAM AND WATER LEAKED FROM THEIR AMSCO 400 STERILIZER. NO REPORT OF INJURY.