inforMED
MalfunctionDXO

COMET II

Received Jan 14, 2026 · Event occurred Nov 6, 2025

Report 2124215-2026-02627 · MDR key 24073134

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

2404-03

Catalog number

2404-03

Lot number

0036902802

Product problems

  • Nonstandard Device
  • Deformation Due to Compressive Stress

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON 17JAN2026. IT WAS REPORTED THAT WIRE WAS KINKED OR DEFORMED. A COMET II PRESSURE GUIDEWIRE WAS SELECTED FOR USE. DURING THE PROCEDURE, THE WIRE WAS KINKED OR DEFORMED LESS THAN 3 CM FROM THE TIP. THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. HOWEVER, DEVICE ANALYSIS REVEALED THAT THE TIP WAS DETACHED FROM THE TIP WELD.

Additional Manufacturer Narrative

DEVICE EVALUATED BY MANUFACTURER. THE COMPLAINT DEVICE WAS RECEIVED FOR PRODUCT ANALYSIS. A VISUAL INSPECTION OF THE DEVICE REVEALED THAT THERE WERE NUMEROUS KINKS THROUGHOUT THE BODY FROM DISTAL TO PROXIMAL, ADDITIONALLY THE TIP WAS FOUND BENT AND DETACHED FROM THE TIP WELD. NO OTHER ISSUES WERE IDENTIFIED DURING THE PRODUCT ANALYSIS. E1-INITIAL REPORTER CITY: (B)(6).