Received Jan 13, 2026 · Event occurred Oct 31, 2025

Report 3008082710-2026-70003 · MDR key 24064178

Device

Generic name

Microwave Dermatologic System

Manufacturer

Miradry Inc.

Model number

MN-MD4000-MC

Catalog number

MN-MD4000-MC

Lot number

20H0603-JP

Product problems

  • Improper or Incorrect Procedure or Method

Patient

30 YR · Female

  • Partial thickness (Second Degree) Burn

Narrative

Additional Manufacturer Narrative

THE MIRADRY SYSTEM LABELING INCLUDES A WARNING TO NOT TREAT THE SAME SITE TWICE IN A GIVEN SESSION. MULTIPLE TREATMENTS TO THE SITE MAY CAUSE SIGNIFICANT DAMAGE TO SUBDERMAL STRUCTURES. TREATMENT DATA REVIEWED CORRESPONDED WITH DUPLICATE TREATMENTS. NO SYSTEM MALFUNCTIONS WERE OBSERVED. REVIEW OF LOT HISTORY RECORDS FOR THE INVOLVED DEVICE CONFIRMED MANUFACTURING STEPS AND PROCESSES WERE MET AND FOLLOWED. PRODUCT MET FINAL INSPECTION AND TESTING REQUIREMENTS PRIOR TO SHIPMENT.

Description of Event or Problem

PATIENT RECEIVED SECOND MIRADRY TREATMENT IN BOTH AXILLA ON (B)(6) 2025, WITH 2 PASSES APPLIED TO ROWS 2 THROUGH 13. ON (B)(6) 2025, PATIENT PRESENTED WITH RIGHT AXILLARY SKIN ULCER SUSPECTED TO BE SECOND-DEGREE BURN. DEBRIDEMENT WITH APPLICATION OF PROSTAGLANDIN OINTMENT AND WOUND DRESSING PERFORMED(B)(6) 2025.