inforMED
MalfunctionJIL

COMBUR 9 TEST

Received Jan 12, 2026 · Event occurred Dec 12, 2025

Report 1823260-2026-00136 · MDR key 24037229

Device

Generic name

Urine Test Strips

Manufacturer

Roche Diagnostics

Catalog number

04510038191

Lot number

86071601

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

NA · Female

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

SECTIONS D9 AND H3 WERE UPDATED. THE CUSTOMER'S TEST STRIPS WERE RECEIVED FOR INVESTIGATION. THE RETENTION AND CUSTOMER MATERIALS SHOWS NO FALSE NEGATIVE RESULTS AND PERFORMED WITHIN SPECIFICATION. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE COMBUR 9 TEST RESULTS FOR 1 PATIENT. THE REPORTER STATED THAT THE PATIENT ALLEGED THAT THEY PURCHASED THE URINE STRIPS FOR HOME URINE DIPSTICK TESTING. THE TEST STRIP ONLY GAVE A POSITIVE RESULT FOR BLOOD, AND WHEN THEY WENT TO THE HOSPITAL 12 HOURS LATER, A URINE DIPSTICK TEST WAS PERFORMED, WHICH WAS POSITIVE FOR LEUKOCYTES, BILIRUBIN, AND BLOOD.

Additional Manufacturer Narrative

THE CUSTOMER'S STRIPS WERE REQUESTED FOR INVESTIGATION. THE INVESTIGATION IS ONGOING.