Received Jan 8, 2026 · Event occurred Dec 10, 2025

Report 8020889-2026-00005 · MDR key 24012515

Device

Generic name

Tube, Bronchial (w/wo Connector)

Model number

125039

Catalog number

125039

Lot number

202505290X

Product problems

  • Gas/Air Leak

Patient

60 YR · Unknown

  • Unintended Extubation

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION: D9, G3, H3 AND H6. H3 EVALUATION SUMMARY: MEDTRONIC CONDUCTED AN INVESTIGATION BASED UPON ALL INFORMATION RECEIVED. THE DEVICE WAS AVAILABLE FOR EVALUATION. VISUAL INSPECTION NOTED A CLEAN-CUT LINEAR SLIT AND A V-SHAPED SLIT ON THE TUBES RESPECTIVELY. AN ATTEMPT MADE TO INFLATE THE RETURNED UNITS FAILED AS THE TRACHEAL CUFFS ON THE TUBES COULD NOT BE INFLATED. IT WAS REPORED THAT THE THREE BRONCO CATHETERS HAD SLOW LEAKS DETECTED IN THE TRACHEAL CUFFS DURING PATIENT INTUBATION ONCE THE TUBE WAS IN SITU, IN A HOSPITAL SETTING. THE REPORTED ISSUE WAS CONFIRMED. THE PRODUCT ANALYSIS NOTED EVIDENCE THAT THE DEVICE WAS NOT USED AS INTENDED. THE MANUFACTURING RECORDS FOR EACH DEVICE ARE THOROUGHLY REVIEWED PRIOR TO RELEASE TO ENSURE THAT IT MEETS ALL MEDTRONIC QUALITY SPECIFICATIONS. A SECONDARY REVIEW OF THE DEVICE HISTORY RECORDS FOUND NO POTENTIALLY CONTRIBUTING FACTORS. THE INSTRUCTIONS INCLUDED WITH THIS DEVICE PROVIDE THE FOLLOWING GUIDANCE: VARIOUS BONY ANATOMICAL STRUCTURES (E.G., TEETH) WITHIN THE INTUBATION ROUTE OR ANY INTUBATION TOOLS WITH SHARP SURFACES PRESENT A THREAT TO MAINTAINING CUFF INTEGRITY. CARE MUST BE TAKEN TO AVOID DAMAGING THE THIN-WALLED CUFFS DURING INTUBATION WHICH WOULD CREATE THE NEED TO SUBJECT THE PATIENT TO THE TRAUMA OF EXTUBATION AND RE-INTUBATION. IF EITHER CUFF IS DAMAGED, THE TUBE SHOULD NOT BE USED. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem

IT WAS REPORTED THAT THE THREE BRONCO CATHETERS HAD SLOW LEAKS DETECTED IN THE TRACHEAL CUFFS DURING PATIENT INTUBATION ONCE THE TUBE WAS IN SITU, IN A HOSPITAL SETTING. THE TUBES WERE REPLACED EVERY TIME IT FAILED AND LASTLY, A 41L TUBE WAS USED, AND SUCCESSFUL INTUBATION WAS ACHIEVED WITH NO FURTHER CUFF ISSUES.

Additional Manufacturer Narrative

D10 CONCOMITANT PRODUCT: 125039, 125039 39FR LT BRONCO CATH PU CUFF X1, LOT# 202505290X. 125039, 125039 39FR LT BRONCO CATH PU CUFF X1, LOT# 202505290X. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.