AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM
Received Jan 7, 2026 · Event occurred Aug 1, 2025
Report 3005099803-2026-00095 · MDR key 23991900
Device
Generic name
Pancreatic Stent, Covered, Metallic, Removable
Manufacturer
Boston Scientific CorporationModel number
M00553520Catalog number
M00553520
Lot number
0036421238
Product problems
- Positioning Failure
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT AN AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM WAS INTENDED TO BE IMPLANTED TO THE PANCREAS TO TREAT A PANCREATIC PSEUDOCYST DURING A PROCEDURE PERFORMED ON (B)(6) 2025. DURING THE PROCEDURE, THE STENT DID NOT RELEASE. THE PROCEDURE WAS COMPLETED, AS THE STENT IS REPORTED TO BE CURRENTLY IMPLANTED. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THIS EVENT. NOTE: THE REPORTED INFORMATION IS CONFLICTING. THE EVENT DESCRIPTION INDICATES THAT THE STENT COULD NOT BE DEPLOYED; HOWEVER, THE STENT WAS ALSO REPORTED AS HAVING BEEN IMPLANTED. DESPITE MULTIPLE GOOD-FAITH ATTEMPTS TO OBTAIN CLARIFICATION, ADDITIONAL INFORMATION TO RESOLVE THIS DISCREPANCY COULD NOT BE OBTAINED.
Additional Manufacturer Narrative
BLOCK H6: IMDRF DEVICE CODE A150201 CAPTURES THE REPORTABLE EVENT OF STENT FAILURE TO DEPLOY. BLOCK H7 AND H9: ADDED TYPE OF REMEDIAL ACTION AND THE CORRECTION/REMOVAL REPORTING # INFORMATION. BLOCK H11: BOSTON SCIENTIFIC IS INITIATING A REMOVAL OF CERTAIN SIZES OF THE AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM (6MM X 8MM, 8MM X 8MM AND 20MM X 10MM) MANUFACTURED SINCE MARCH 1, 2025. THIS ACTION IS BEING TAKEN DUE TO INCREASED REPORTS OF STENT DEPLOYMENT AND EXPANSION ISSUES WITH THESE CONFIGURATIONS. THESE ISSUES ONLY OCCUR AT THE TIME OF STENT DELIVERY AND ARE EXPECTED TO BE NOTICED BY THE PHYSICIAN. PATIENTS WHO HAVE BEEN TREATED WITH A SUCCESSFULLY IMPLANTED AXIOS STENT SHOULD CONTINUE TO FOLLOW STANDARD OF CARE AND ARE NOT AFFECTED BY THIS ISSUE. A LETTER WAS SENT OUT TO MATERIALS MANAGERS/HEALTH CARE PROFESSIONALS ON 19-DEC-2025 TO IMMEDIATELY STOP FURTHER DISTRIBUTION OR USE OF ANY AFFECTED 6MM X 8MM, 8MM X 8MM AND 20MM X 10MM AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM. THE LETTER INDICATES TO REMOVE THESE DEVICES FROM INVENTORY AND SEGREGATE THEM IN A SECURE LOCATION UNTIL THEY CAN BE RETURNED TO BOSTON SCIENTIFIC. THE REFERENCED AXIOS STENT AND ELECTROCAUTERY ENHANCED DELIVERY SYSTEM MET ALL SPECIFICATIONS PRIOR TO FINAL APPROVAL FOR DISTRIBUTIONS/SALE.
Remedial action
- Recall