MalfunctionDXO

TRUWAVE DISPOSABLE PRESSURE TRANSDUCER

Received Jan 7, 2026 · Event occurred Aug 28, 2025

Report 2015691-2026-10219 · MDR key 23988187

Device

Generic name

Transducer, Pressure, Catheter Tip

Model number

PX260

Lot number

66273564

Product problems

  • Contamination

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDED INFORMATION TO SECTION H6 (TYPE OF INVESTIGATION). UPDATED SECTION H6 (COMPONENT CODE), H6 (INVESTIGATION FINDINGS) AND H6 (INVESTIGATIONS CONCLUSIONS). THE MANUFACTURING RECORDS WERE REVIEWED FOR THE LOT INVOLVED AND THERE IS NO INDICATION OF A RELATED NONCONFORMANCE. ALL PROCESS PARAMETERS WERE MET AND INSPECTIONS PASSED SUCCESSFULLY. IT WAS DETERMINED THAT THIS ISSUE IS RELATED TO THE SUPPLIER PROCESS; THEREFORE, THE SUPPLIER HAS BEEN NOTIFIED IN ORDER TO PERFORM AN INVESTIGATION TO AVOID THE RECURRENCE OF THIS ISSUE. THERE ARE MANUFACTURING CONTROLS FOR THE RELATED MALFUNCTION.

Additional Manufacturer Narrative

THE DEVICE INVOLVED IN THIS COMPLAINT WAS RECEIVED BY OUR PRODUCT EVALUATION LABORATORY FOR A FULL EXAMINATION. THE REPORT OF BROKEN CONNECTOR COULD NOT BE CONFIRMED. AS RECEIVED, ALL CONNECTIONS WERE TIGHT AND SECURE. NO LEAKAGE WAS DETECTED FROM THE KIT DURING LEAK TEST. HOWEVER, FILTER WAS FOUND DISLODGED INSIDE DRIP CHAMBER DURING VISUAL EXAMINATION. NO OTHER VISIBLE DAMAGE OR INCONSISTENCY WAS OBSERVED FROM THE KIT. FURTHER ENGINEERING INVESTIGATION IS BEING PERFORMED, WHEN THE PRODUCT INVESTIGATION IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE SENT WITH THE INVESTIGATION RESULTS.

Description of Event or Problem

AS REPORTED, THIS PRESSURE MONITORING SET HAD THE CONNECTOR BROKEN. THERE IS NO ALLEGATION OF PATIENT INJURY. AS PER PRODUCT EVALUATION FINDINGS, FILTER WAS FOUND DISLODGED INSIDE DRIP CHAMBER DURING VISUAL EXAMINATION. PATIENT DEMOGRAPHICS WERE NOT PROVIDED.