RIGIDFIX FEM 3.3MM S/T XPIN
Received Jan 6, 2026 · Event occurred Dec 16, 2025
Report 1221934-2026-00070 · MDR key 23977558
Device
Generic name
Pin, Fixation, Smooth
Manufacturer
Depuy Mitek Llc UsCatalog number
210133
Lot number
213L913
Product problems
- Break
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: D9, H3, H6: THE ACTUAL DEVICE HAS BEEN RETURNED AND IS CURRENTLY PENDING EVALUATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, AND IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL MEDWATCH REPORT WILL BE SUBMITTED ACCORDINGLY.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: UPDATE: D4, H4: EXPIRATION DATE, PRIMARY UDI NUMBER AND DEVICE MANUFACTURE DATE HAS BEEN ADDED. INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO DEPUY SYNTHES FOR EVALUATION. VISUAL INSPECTION OF THE RETURNED DEVICE FOUND THAT THE DEVICE COMES IN USED CONDITION. THE HOLE KIT WAS RETURNED. THE TROCAR AND 2 SHEATHS ARE IN GOOD CONDITION, NO STRUCTURAL ANOMALIES WERE FOUND. UPON REVIEWING THE 2 TRANSPARENT CROSS PINS, IT WAS FOUND THAT THEY ARE BENT. NO OTHER ANOMALIES WERE FOUND. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE RIGIDFIX FEM 3.3MM S/T XPIN WOULD HAVE NOT CONTRIBUTED TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, THE POTENTIAL CAUSE FOR THE BENT CROSS PINS IS TRACED TO THERMAL DEGRADATION OF THE POLYMERS DUE TO AN EXPOSURE TO HIGHER TEMPERATURES THAN THE RECOMMENDED WHILE IN STOCK. AS PER INSTRUCTIONS FOR USE (IFU), STORE IN A COOL DRY AREA (BELOW 26°C, 80°F). IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY RECORD REVIEW: A DEVICE HISTORY REVIEW WAS PERFORMED AND NO NON-CONFORMANCES WERE IDENTIFIED RELATED TO THE REPORTED CONDITION.
Additional Manufacturer Narrative
THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE. H11: ADDITIONAL NARRATIVE: E1: THE REPORTER¿S COMPLETE FACILITY ADDRESS WAS NOT PROVIDED. AS OF THIS DATE, THE DEVICE HAS NOT BEEN RETURNED FOR EVALUATION; THEREFORE, THE REPORTED CONDITION CANNOT BE CONFIRMED AND/OR DUPLICATED.
Description of Event or Problem
THIS IS REPORT 1 OF 3 FOR (B)(4). IT WAS REPORTED THAT DURING A KNEE SURGERY, IT WAS DISCOVERED THAT THE ANCHOR ON THE RIGIDFIX FEM 3.3MM S/T XPIN DEVICE BROKE OFF. ALL THE BROKEN PARTS WERE REMOVED. A REPLACEMENT DEVICE WAS USED TO CONTINUE THE SURGERY, BUT THE SAME ISSUE OCCURRED AGAIN. FURTHERMORE, THE NEEDLE ON THE 5.5 HEALIX ADVANCE KNTLSS BR ALSO BROKE OFF. ANOTHER ANCHOR AND NEEDLE WERE USED TO COMPLETE THE SURGERY. IT WAS NOT REPORTED IF THERE WERE ANY DELAYS TO THE SURGICAL PROCEDURE. IT WAS REPORTED THAT A SPARE DEVICE WAS USED TO COMPLETE THE SURGERY. THERE WERE NO REPORTS OF INJURIES, MEDICAL INTERVENTION OR PROLONGED HOSPITALIZATION. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED.