inforMED
OtherMOS

SIEMENS 3.0T TXRX 15 CH KNEE

Received Dec 20, 2011 · Event occurred Nov 16, 2011

Report 3007350713-2011-00002 · MDR key 2396550

Device

Generic name

Coil, Magnetic Resonance, Specialty

Model number

Q7000008

Catalog number

QED000008

Lot number

NA

Product problems

  • Adverse Event Without Identified Device or Use Problem

Patient

UNK

  • Superficial (First Degree) Burn

Narrative

Description of Event or Problem

CUSTOMER HAD BURNS ON KNEE (RED IRRITATED LOOK). NO OTHER ISSUES ON COIL DETECTED, BUT COIL EXCHANGED FOR SAFETY CONCERNS.

Additional Manufacturer Narrative

INVESTIGATION DETAILS: UPON RECEIVING THE COIL, NORMAL COIL CONFIDENCE TESTING WAS PERFORMED AND NO FAULT WAS FOUND. FURTHER INVESTIGATION FOUND THAT THE TUNING OF SEVERAL RECEIVE CHANNELS WAS OUTSIDE THE DATA SHEET SPECIFICATION. THE SOURCE OF THE ELEMENTS BEING OFF-TUNED WAS LINKED TO THE VARIABLE CAPACITORS ON EACH COIL ELEMENT. HOWEVER, BECAUSE THESE VARIABLE CAPACITORS ARE ALSO USED IN THE DECOUPLING CIRCUITS, WE PERFORMED TESTING TO CONFIRM AN OFF-TUNED DECOUPLING CIRCUIT DID NOT CONTRIBUTE TO THE REPORTED INCIDENT. SUMMARY: BASED ON THE TESTING PERFORMED, WE CONCLUDED THAT THE OFF-TUNED DECOUPLING CIRCUITS DID NOT CONTRIBUTE TO THE REPORTED INCIDENT.