ELECSYS BRAHMS PCT

Received Jan 2, 2026 · Event occurred Dec 5, 2025

Report 1823260-2026-00002 · MDR key 23942961

Device

Generic name

Procalcitonin Assay

Manufacturer

Roche Diagnostics

Catalog number

08828644190

Lot number

864044

Product problems

  • Incorrect Measurement
  • High Test Results
  • Low Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE COBAS 6000 E601 MODULE SERIAL NUMBER WAS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS BRAHMS PCT RESULTS FROM THE COBAS 6000 E601 MODULE. THE INITIAL RESULT WAS >100 NG/ML. THE PHYSICIAN REQUESTED MANUAL DILUTION, AND THE CUSTOMER PERFORMED A 1:4 MANUAL DILUTION. THE RESULT WAS 15 NG/ML (X4=60 NG/ML). THE CUSTOMER PERFORMED ANOTHER 1:2 MANUAL DILUTION OF THE ORIGINAL SAMPLE, PRODUCING A RESULT OF 52 NG/ML. THE CUSTOMER DID NOT KNOW WHICH RESULT WAS CORRECT.

Additional Manufacturer Narrative

THE INVESTIGATION WAS UNABLE TO DETERMINE A SPECIFIC ROOT CAUSE. A GENERAL REAGENT OR ANALYZER PROBLEM WAS NOT PRESENT, BECAUSE THE QC PRIOR TO THE EVENT WAS WITHIN RANGES.