Received Dec 30, 2025 · Event occurred Dec 17, 2025

Report 3007417753-2025-000000 · MDR key 23919755

Device

Generic name

Table And Attachments, Operating-room

Model number

111860

Catalog number

1118.60

Product problems

  • Unintended Movement
  • Unintended Movement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ON DEC. 17TH, 2025, MAQUET SUZHOU BECAME AWARE OF AN UNEXPECTED DESCENT OF A 1118.60 LEG PLATE DURING SURGERY. THIS WAS CAUSED BY A SUDDEN OIL LEAK FROM THE PNEUMATIC CYLINDER, WHICH REDUCED THE FORCE HOLDING THE LEG PLATE IN POSITION. THERE WAS NO INJURY REPORTED.

Additional Manufacturer Narrative

(B)(4). GETINGE FIELD SERVICE ENGINEER VISITED THE HOSPITAL AND PERFORMED THE INSPECTION ON THE INVOLVED UNIT. AN OIL LEAKAGE WAS OBSERVED ON THE PNEUMATIC CYLINDER OF LEG PLATE. GETINGE FIELD SERVICE ENGINEER REPLACED THE DEFECTIVE PNEUMATIC CYLINDER WITH NEW ONE, THE UNIT WAS VERIFIED IN GOOD CONDITION AND COULD PUT IN USE. ADDITIONAL INFORMATION WILL BE PROVIDED FOLLOWING THE CONCLUSION OF THE INVESTIGATION.