ELECSYS ACTH
Received Dec 29, 2025 · Event occurred Dec 9, 2025
Report 1823260-2025-05351 · MDR key 23905893
Device
Generic name
Adrenocorticotropic Hormone (acth) Test System
Manufacturer
Roche DiagnosticsCatalog number
07026684190
Lot number
ASKU
Product problems
- High Test Results
- High Test Results
Patient
65 YR · Female
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE HIGH RESULTS FOR 1 PATIENT TESTED FOR ELECSYS ACTH (ACTH) ON A COBAS E 801 ANALYTICAL UNIT. THE ACTH RESULTS ARE ALLEGEDLY HIGH WITH NO KNOWN CAUSE, AS THE PATIENT IS ASYMPTOMATIC (PREVIOUSLY CUSHING'S DISEASE). THE INITIAL RESULT WAS 118,000 PG/ML. THE SAMPLE WAS REPEATED WITH VARIOUS DILUTIONS: X5: > 2000 PG/ML WITH A DATA FLAG, X10: >2000 PG/ML WITH A DATA FLAG, X20: >2000 PG/ML WITH A DATA FLAG, X40: >2000 PG/ML WITH A DATA FLAG, X75: 118,350 PG/ML, X100: 126,400 PG/ML. ON (B)(6) 2025, A NEW SAMPLE WAS OBTAINED, AND THE RESULT WAS APPROXIMATELY 25,000 PG/ML.
Additional Manufacturer Narrative
THE E801 ANALYZER SERIAL NUMBER WAS (B)(6). SAMPLE MATERIAL FROM THE PATIENT WAS RECEIVED FOR INVESTIGATION. THE CUSTOMER'S HIGH ACTH RESULTS WERE REPRODUCED AT THE INVESTIGATION SITE. SAMPLE MATERIAL FROM THE PATIENT WAS SENT TO AN EXTERNAL LABORATORY AND TESTED BY THE SIEMENS METHOD, WHERE THE ACTH WAS ALSO EXTREMELY HIGH (80,163 PG/ML). INTERFERENCE TESTING USING SIZE EXCLUSION CHROMATOGRAPHY (SEC) WAS PERFORMED. NO INTERFERENCE WAS FOUND. PEPTIDE MAPPING ANALYSIS OF THE SAMPLE FRACTIONS SUGGESTS THE HIGH ELECSYS ACTH RESULTS ARE CONSISTENT WITH AN ERRONEOUS DETECTION OF A DIFFERENT PEPTIDE FROM THE HIGHER-LEVEL PROOPIOMELANOCORTIN (POMC) PROTEIN.