MalfunctionFMZ
GIRAFFE INCUBATOR CARESTATION
Received Dec 26, 2025 · Event occurred Dec 1, 2025
Report 2112667-2025-10030 · MDR key 23898589
Device
Generic name
Incubator, Neonatal
Manufacturer
Datex-ohmeda, Inc.Model number
CS1Catalog number
NA
Lot number
NA
Product problems
- No Flow
- No Flow
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE SERVO O2 BOARD WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF AUXILIARY OXYGEN. THERE WAS NO PATIENT INVOLVEMENT.