inforMED
MalfunctionFMZ

GIRAFFE INCUBATOR CARESTATION

Received Dec 26, 2025 · Event occurred Dec 1, 2025

Report 2112667-2025-10028 · MDR key 23898576

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Model number

CS1

Catalog number

NA

Lot number

NA

Product problems

  • No Flow
  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF AUXILIARY OXYGEN. THERE WAS NO PATIENT INVOLVEMENT.

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE SERVO O2 BOARD WAS REPLACED TO RESOLVE THE ISSUE. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: HCS RESEARCH PARK - 9900 INNOVATION DRIVE USA WAUWATOSA, WI 53226.