inforMED
MalfunctionDTL

INDEFLATOR

Received Dec 23, 2025 · Event occurred Dec 2, 2025

Report 2024168-2025-12221 · MDR key 23878200

Device

Generic name

Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Model number

1003327

Catalog number

1003327

Lot number

60658754

Product problems

  • Leak/Splash
  • Leak/Splash

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED IN THE POSTERIOR DESCENDING ARTERY (PDA) WITH MODERATE CALCIFICATION AND MODERATE TORTUOSITY. THE INDEFLATOR LOSES PRESSURE AND DOES NOT INFLATE THE BALLOON/STENT. ANOTHER INDEFLATOR WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.

Additional Manufacturer Narrative

THE DEVICE WAS RETURNED FOR ANALYSIS. THE REPORTED LEAK WAS UNABLE TO BE CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTATIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. ADDITIONALLY, A QUERY OF THE COMPLAINT HANDLING DATABASE FOR THE REPORTED LOT REVEALED THERE IS NO INDICATION OF A LOT SPECIFIC ISSUE. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, A DEFINITIVE CAUSE FOR THE REPORTED DIFFICULTIES COULD NOT BE DETERMINED. AS THE REPORTED LEAK WAS UNABLE TO BE CONFIRMED DURING RETURN ANALYSIS, IT IS POSSIBLE THAT THE CONNECTION PORT WAS NOT FULLY CONNECTED OR TIGHTENED RESULTING IN THE REPORTED LEAK; HOWEVER, THIS CANNOT BE CONFIRMED. THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE.

Additional Manufacturer Narrative

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT.

Description of Event or Problem

IT WAS REPORTED THAT THE PROCEDURE WAS PERFORMED IN THE POSTERIOR DESCENDING ARTERY (PDA) WITH MODERATE CALCIFICATION AND MODERATE TORTUOSITY. THE INDEFLATOR LOSES PRESSURE AND DOES NOT INFLATE THE BALLOON/STENT. ANOTHER INDEFLATOR WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT AND THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE. NO ADDITIONAL INFORMATION WAS PROVIDED.