MalfunctionFMZ
GIRAFFE INCUBATOR
Received Dec 23, 2025 · Event occurred Nov 25, 2025
Report 2112667-2025-09945 · MDR key 23877674
Device
Product problems
- No Flow
- No Flow
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
GE HEALTHCARE'S INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.
Description of Event or Problem
IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN OXYGEN FAILURE. THERE WAS NO PATIENT INVOLVEMENT.
Additional Manufacturer Narrative
A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 SERVER COOLING FAN WAS REPLACED TO RESOLVE THE ISSUE.