inforMED
MalfunctionMUJ

LEKSELL GAMMAPLAN

Received Dec 23, 2025

Report 3015232217-2025-00063 · MDR key 23877055

Device

Generic name

Radionuclide Rt Treatment Planning System

Lot number

11.4.1

Product problems

  • Improper or Incorrect Procedure or Method
  • Improper or Incorrect Procedure or Method

Patient

NA · Unknown

  • Unintended Radiation Exposure
  • Unintended Radiation Exposure

Narrative

Additional Manufacturer Narrative

DUE TO AN ADMINISTRATIVE ERROR, THE REPORT WAS ORIGINALLY SUBMITTED UNDER ., WHICH WAS AN INCORRECT/INCOMPLETE REFERENCE NUMBER. THE CORRECT REFERENCE SHOULD HAVE BEEN .. THEREFORE, THIS REPORT HAS BEEN SUBMITTED AS AN INITIAL AND FINAL TO CORRECT THE REFERENCE NUMBER. H6 UPDATED. H11 UPDATED: THE INVESTIGATION WAS COMPLETED BY CONDUCTING A THOROUGH EVALUATION OF THE PRODUCT AND THE REPORTED INFORMATION. THE CUSTOMER REPORTED THAT AN OBSOLETE STEREOTACTIC REFERENCE WAS USED FOR TREATMENT. IT WAS ASCERTAINED THAT A TREATMENT PLAN WAS CREATED USING FRAMELESS MRI IMAGES, G-FRAME FIXATION, AND CONE BEAM COMPUTED TOMOGRAPHY (CBCT) AS THE STEREOTACTIC REFERENCE. DURING THE PLANNING PROCESS, THE CUSTOMER ELECTED TO REPOSITION THE FRAME AND ACQUIRED A NEW CBCT. HOWEVER, DUE TO A USE ERROR, THE TREATMENT PLAN WAS NOT CORRECTLY UPDATED TO REFLECT THE NEW FRAME POSITION. AS A RESULT, THE TREATMENT WAS DELIVERED BASED ON THE PREVIOUS FRAME PLACEMENT. THE PLAN HAD BEEN APPROVED AND PARTIALLY DELIVERED BEFORE THE OPERATOR IDENTIFIED THE DISCREPANCY AND PAUSED THE TREATMENT. THE SYSTEM WORKED AS DESIGNED AND INTENDED. ELEKTA HAVE NOT BEEN PROVIDED WITH INFORMATION REGARDING THE PATIENT'S CONDITION. AN IMPORTANT FIELD SAFETY NOTICE (100-01-102-019) WILL BE SENT TO ALL AFFECTED CUSTOMERS FROM 17 DECEMBER 2025 (ELEKTA REFERENCE #: (B)(4)).

Description of Event or Problem

THE CUSTOMER REPORTED THAT AN OBSOLETE STEREOTACTIC REFERENCE USED FOR TREATMENT.