CADD LEGACY PCA PUMPS
Received Dec 23, 2025 · Event occurred Dec 1, 2025
Report 3012307300-2025-13671 · MDR key 23876503
Device
Generic name
Pump, Infusion, Pca
Manufacturer
Icu Medical IncModel number
6300Catalog number
21-6300-09JP
Product problems
- Device Alarm System
- Device Alarm System
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
H3/H6: INVESTIGATION INCLUDING ROOT CAUSE ANALYSIS IS IN PROGRESS. A SUPPLEMENTAL MDR WILL BE FILED AS NECESSARY IN ACCORDANCE WITH 21 CFR 803.56 WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE.
Description of Event or Problem
IT WAS REPORTED THAT THE ALARM IS CONTINUOUSLY SOUNDING AND DOES NOT DEACTIVATE. THE EVENT OCCURRED WHILE PATIENT USE AT THE FACILITY. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED.
Additional Manufacturer Narrative
H3: THE DEVICE WAS RECEIVED FOR EVALUATION. ICU MEDICAL JAPAN HAS RECEIVED ONE REPAIR REQUEST IN THE PAST YEAR. THE MOST RECENT REPAIR REQUEST WAS MADE ON 2025-JAN-20 (RO#1401592). VISUAL AND FUNCTIONAL TESTS WERE PERFORMED, WHICH FOUND THE ROOT CAUSE OF THE EVENT WAS AN ANOMALY IN THE GEAR OF THE MOTOR. THE DEVICE WAS RETURNED TO THE CUSTOMER WITH NO REPAIR PROVIDED AT THEIR REQUEST.