NATURA
Received Dec 23, 2025
Report 9618003-2025-03393 · MDR key 23875466
Device
Generic name
Protector, Ostomy
Manufacturer
Convatec Dominican Republic IncModel number
404592Lot number
4L05864
Product problems
- Malposition of Device
- Malposition of Device
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
DEVICE 4 OF 16. BASED ON THE AVAILABLE INFORMATION, THIS EVENT IS DEEMED TO BE A REPORTABLE MALFUNCTION. PHOTOGRAPH, VIDEO AND/OR PHYSICAL SAMPLE EVALUATION: NO PHOTOGRAPH ASSOCIATED WITH THIS CASE WAS RECEIVED. NO RETURN SAMPLE HAS BEEN RECEIVED FOR THIS COMPLAINT. BATCH RECORD REVISION RESULTS: LOT 4L05864 WAS MANUFACTURED ON 28/NOV/2024, IN CONVEX 2 PC LINE, WITH A TOTAL OF (B)(4) MARKET UNITS (MKUS). THE COMPLAINT INVESTIGATOR PERFORMED A BATCH RECORD REVIEW ON 12/NOV/2025, TO VERIFY IF ALL THE APPLICABLE PROCEDURES WERE FOLLOWED AND NO ISSUES WERE FOUND; ALL THE COMPONENTS FOR ASSEMBLY WERE CORRECT PER BILL OF MATERIALS (BOM) AND ALL THE TOOLING INFORMATION DOCUMENTED WAS ALSO CORRECT, SYSTEM APPLICATION PRODUCT (SAP) MATERIAL IDENTIFICATION (ID) 1156500 AND MANUFACTURING ORDER 1790416. REVIEW OF THE BATCH RECORD SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCY RELATED TO THIS ISSUE WERE FOUND WITHIN THE DOCUMENTATION. HISTORICAL COMPLAINTS REVIEW: ON 12/NOV/2025, COMPLAINT INVESTIGATOR RAN A QUERY IN DATABASE IN ORDER TO VERIFY THE COMPLAINTS REPORTED FOR 4L05864 LOT FOR THE MALFUNCTION ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (MOLDABLE ONLY)¿ AND AS RESULT NO ADDITIONAL TYPE 2 COMPLAINTS WERE IDENTIFIED DURING THIS SEARCH AS PER WORK INSTRUCTIONS (WI). HISTORICAL NONCONFORMANCE REVIEW: ON 12/NOV/2025, COMPLAINT INVESTIGATOR RAN A QUERY IN DATABASE LOOKING FOR ANY IN PROCESS NONCONFORMANCE / CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) (S) ASSOCIATED TO THE MALFUNCTION ¿SKIN BARRIER STARTER HOLE IS DEFECTIVE (E.G MISALIGNMENT OR OFF CENTER), LEAKAGE MAY OCCUR (MOLDABLE ONLY)¿ FOR THE LOT NUMBER 4L05864 AND AS RESULT, NO NONCONFORMANCE / CAPA (S) FOR THIS MALFUNCTION WERE GENERATED DURING THE MANUFACTURING PROCESS OF THE REFERENCED LOT. CURRENT QUALITY CONTROLS: BASED ON THE PROCESS INSTRUCTION (PI), THE FOLLOWING TESTS ARE PERFORMED IN THE MANUFACTURING LINES, IN ORDER TO IDENTIFY THIS FAILURE MODE IN OUR MANUFACTURING PROCESS: ADHESIVE TO CONVEX DISC CENTERING (VISUAL): NOT LESS THAN (NLT) 0.095" (2.4MM) FROM IDENTIFICATION DOCUMENT (ID) OF THE ADHESIVE CENTER HOLE TO ID OF THE CONVEX DISC AT ANY POINT. (QUALITY CONTROL (QC) TOOL) PICTURE FRAME SECONDARY REFERENCE PRODUCT (SRP): NLT 0.504" (12.8MM), WIDTH OF THE SRP LEFT AT ANY POINT ON THE PICTURE FRAME. (QUALITY CONTROL (QC) TOOL) PICTURE FRAME SRP TO CONVEX DISC CENTERING: NLT 0.040" (1.0MM) FROM THE OUTER DIAMETER OF THERMOFORMED DISC TO COLLAR SRP KISS CUT INNER DIAMETER. (QUALITY CONTROL (QC) TOOL) FREQUENCY: HOURLY. SAMPLE QUANTITY: 5 SAMPLES.. ACCEPTANCE CRITERIA: ACCEPT = 0 / REJECT = 1. DEFECT RATE ANALYSIS: THERE HAVE ONLY BEEN 16 DEFECTIVE PARTS CONFIRMED TO DATE FROM A LOT SIZE OF (B)(4) PRODUCTS. THIS REPRESENTS A DEFECT RATE OF ONLY 0.04% WHICH IS WELL WITHIN AN APPROPRIATE ACCEPTABLE QUALITY LEVEL (AQL) FOR LEAKAGE WHICH SHOULD BE 2.5% BASED ON OUR PROCESS INSTRUCTION (PI). IN ADDITION, ALL THE IN-PROCESS TESTING ON THIS LOT DID NOT FIND A SINGLE DEFECTIVE UNIT, WHICH CONFIRMS THAT THE LOT IS UNLIKELY TO BREACH AN AQL OF 2.5 THIS ISSUE CERTAINLY APPEARS TO BE AN ISOLATED INCIDENT, BUT MORE IMPORTANTLY TO DATE, IT IS WELL WITHIN OUR ACCEPTED AQL LEVEL FOR THIS TYPE OF FAILURE MODE OR DEFECT. CONCLUSIONS: A REVIEW OF BATCH RECORD WAS COMPLETED AND SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE MANUFACTURING PROCESS AND FINAL PRODUCT RELEASE HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DISCREPANCIES RELATED TO THIS ISSUE WERE FOUND WITHIN THE DOCUMENTATION. ADDITIONALLY, A REVIEW OF HISTORICAL COMPLAINTS WAS PERFORMED, AND NO ADDITIONAL COMPLAINTS WERE IDENTIFIED, AND A REVIEW OF HISTORICAL NONCONFORMANCES SHOWED NO ADDITIONAL NONCONFORMANCES. THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET PRODUCT MONITORING REVIEW PROCESS, STANDARD OPERATING PROCEDURE (SOP). TO DATE NO ADDITIONAL INFORMATION HAS BEEN RECEIVED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. FDA REGISTRATION NUMBER. REPORTING SITE: 1049092. MANUFACTURING SITE: 9618003.
Description of Event or Problem
THE END USER REPORTED THAT SIXTEEN WAFERS FROM TWO BOXES WITH SAME LOT NUMBER HAD OFF CENTERED STARTER HOLE. THE PRODUCT WAS NOT USED. NO PHOTO WAS AVAILABLE AT THAT TIME.