PRO+
Received Dec 19, 2025 · Event occurred Dec 4, 2025
Report 3006697241-2025-00344 · MDR key 23854854
Device
Generic name
Bed, Flotation Therapy, Powered
Manufacturer
Baxter Healthcare CorporationModel number
P7924A03Catalog number
P7924A03
Lot number
N/A
Product problems
- Material Split, Cut or Torn
- Material Split, Cut or Torn
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE BAXTER TECHNICIAN FOUND THE POWER CORD NEEDED TO BE REPLACED. IT IS NECESSARY FOR THE PRO+ MATTRESS TO HAVE AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU DO ANNUAL PREVENTIVE MAINTENANCE (PM) AND TESTING FOR JOINT COMMISSION CERTIFICATION. EXAMINE THE ENTIRE LENGTH OF THE MATTRESS POWER CORD. MAKE SURE THERE ARE NO CUTS OR EXPOSED WIRES. MAKE SURE THE PLUG IS A ONE-PIECE MOLDED PLUG ASSEMBLY. EXAMINE THE PLUG FOR DAMAGE. REPLACE THE MATTRESS POWER CORD IF THE PLUG SHOWS: DISCOLORATION OF THE PLUG MOLDING ON OR AROUND THE PLUG BLADES. SIGNS OF CRACKING. LOOSE FIT OF THE PLUG BLADE (THE PLUG BLADE MOVES IN THE MOLDING). A SEARCH OF THE BAXTER MAINTENANCE RECORDS DID NOT SHOW BAXTER PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNKNOWN IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE TECHNICIAN REPLACED THE POWER CORD TO RESOLVE THE REPORTED EVENT. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
Description of Event or Problem
ON 04-DEC-2025, A COMPANY REPRESENTATIVE - SERVICE PERSONNEL CONTACTED TECHNICAL SERVICE TO REPORT THAT PRO+ SURFACE 36" WIDE W/O XRAY (PRODUCT CODE: P7924A03, SERIAL NUMBER: (B)(6), HAD POWER CORD DAMAGED WITH COPPER WIRES EXPOSED. THE PROCESS STEP DURING WHICH THIS OCCURRED WAS UNKNOWN. THERE WAS NO PATIENT/USER INJURY, MEDICAL INTERVENTION, SYMPTOM, OR ADVERSE EVENT REPORTED.