BEACH CHAIR H

Received Dec 19, 2025 · Event occurred Nov 20, 2025

Report 3006697241-2025-00342 · MDR key 23853523

Device

Generic name

Support, Patient Position

Model number

2077600

Catalog number

2077600

Lot number

N/A

Product problems

  • Fracture
  • Fracture

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE BEACH CHAIR SHOULDER POSITIONER IS DESIGNED TO POSITION AND SUPPORT THE PATIENT¿S HEAD AND TORSO IN A VARIETY OF SURGICAL PROCEDURES INCLUDING, BUT NOT LIMITED TO ORTHOPEDIC SURGERIES LIKE, ROTATOR CUFF, TOTAL SHOULDER, REVERSE TOTAL SHOULDER, SLAP REPAIR AND OTHER SHOULDER SURGERY. THESE DEVICES ARE INTENDED TO BE USED BY HEALTHCARE PROFESSIONALS WITHIN THE OPERATING ROOM SETTING. THE BEACH CHAIR WAS RECEIVED BY BAXTER. UPON INSPECTION, IT WAS DETERMINED THAT THE BACKBOARD WAS CRACKED IN THE UPPER PORTION. BACKBOARD, SCREW COVERS AND VELCRO WERE REPLACED. BASED ON THIS INFORMATION, NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME. ALTHOUGH THERE WAS NO REPORTED INJURY WITH THIS EVENT, IF THE REPORT OF A BEACH CHAIR BACKBOARD BEING CRACKED WERE TO RECUR, IT COULD POTENTIALLY CAUSE SERIOUS INJURY OR DEATH. THEREFORE BAXTER IS REPORTING THIS EVENT.

Description of Event or Problem

BAXTER RECEIVED A REPORT STATING THAT BEACH CHAIR HAD A CRACK IN THE FIBERGLASS BACK SECTION WHEN A PATIENT WAS POSITIONED. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.