Received Dec 18, 2025 · Event occurred Nov 21, 2025

Report 0002936485-2025-00998 · MDR key 23843440

Device

Generic name

Cart, Emergency, Cardiopulmonary (excluding Equipment)

Catalog number

0240099155

Product problems

  • No Display/Image
  • Electrical Shorting
  • No Display/Image
  • Electrical Shorting

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

ADDITIONAL INFORMATION WILL BE PROVIDED ONCE THE INVESTIGATION HAS BEEN COMPLETED. THE DEVICE MANUFACTURER DATE IS NOT KNOWN AT THIS TIME. HOWEVER, SHOULD IT BECOME AVAILABLE IT WILL BE PROVIDED IN FUTURE REPORTS.

Description of Event or Problem

IT WAS REPORTED THAT THE CART SHUT OFF MID CASE. NO PATIENT HARM OR SURGICAL DELAY. THEREFORE, THERE WAS LOSS OF IMAGE.

Additional Manufacturer Narrative

THREE ATTEMPTS HAVE BEEN MADE TO RETRIEVE THE DEVICE FOR EVALUATION, AND IT IS YET TO BE RECEIVED IN-HOUSE. THIS COMPLAINT INVESTIGATION WAS CLOSED BASED ON THE DEVICE NOT RECEIVED, THEREFORE THE REPORTED FAILURE MODE WAS NOT CONFIRMED. IN THE EVENT THAT THE DEVICE IS RECEIVED, THE COMPLAINT WILL BE REOPENED AND THE INVESTIGATION WILL BE UPDATED WITH NEW RESULTS. ALLEGED FAILURE: CART SHUT OFF MID-CASE. PROBABLE ROOT CAUSE: OVERLOADING THE CART (EXCESSIVE CURRENT DRAW). USE ERROR. THE REPORTED FAILURE MODE WILL BE MONITORED FOR FUTURE REOCCURRENCE.

Description of Event or Problem

IT WAS REPORTED THAT THE CART SHUT OFF MID CASE. NO PATIENT HARM OR SURGICAL DELAY. THEREFORE, THERE WAS LOSS OF IMAGE.