BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM HEPARIN
Received Dec 17, 2025 · Event occurred Nov 20, 2025
Report 1917413-2025-01690 · MDR key 23835446
Device
Generic name
Lymphocyte Separation Medium
Manufacturer
Becton Dickinson & Co (franklin Lakes)Catalog number
362753
Lot number
5196666
Product problems
- Short Fill
- Short Fill
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
IT WAS REPORTED WHILE USING BD VACUTAINER® CPT¿ CELL PREPARATION TUBE WITH SODIUM HEPARIN, AN UNSPECIFIED NUMBER OF TUBES HAD VACUUM LOSS RESULTING IN UNDERFILLED TUBES. THE TUBE WAS REPLACED. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Additional Manufacturer Narrative
INVESTIGATION SUMMARY: BD HAD NOT RECEIVED SAMPLES OR PHOTOS FOR INVESTIGATION. THEREFORE, 59 RETENTION SAMPLES FROM BD INVENTORY WERE VISUALLY INSPECTED AND NO ISSUES WERE OBSERVED; ALL STOPPERS WERE CORRECTLY PLACED ON THE TUBES. ADDITIONALLY, 10 RETAIN SAMPLES WERE FUNCTIONALLY TESTED FOR DRAW VOLUME AND ALL TUBES TESTED WITHIN SPECIFICATION REQUIREMENTS. BASED ON A REVIEW OF THE DEVICE HISTORY RECORD FOR THE INCIDENT LOT, ALL PRODUCT SPECIFICATIONS AND REQUIREMENTS FOR LOT RELEASE WERE MET. THERE WERE NO RELATED QUALITY ISSUES DURING MANUFACTURING OF THE PRODUCT. THIS COMPLAINT IS UNABLE TO BE CONFIRMED FOR THE INDICATED FAILURE MODE UNDERFILL/VACUUM ISSUE. BD WAS NOT ABLE TO IDENTIFY A ROOT CAUSE FOR THE INDICATED FAILURE MODE. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. OUR BUSINESS TEAM REGULARLY REVIEWS THE COLLECTED DATA FOR IDENTIFICATION OF EMERGING TRENDS.