Received Dec 17, 2025 · Event occurred Dec 10, 2025

Report MW5180704 · MDR key 23829954

Device

Generic name

Emergency Response Safety Kit

Model number

36-23446, REV3

Product problems

  • Application Program Problem: Dose Calculation Error

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

INFORMATION CONTAINED ON THE AIRLIFE BROSELOW PEDIATRIC EMERGENCY REFERENCE TAPE FOR 2025 EDITION, 36-23446, REV3. HAS AN INCORRECT CALCULATION. IN THE ORANGE WEIGHT SECTION, UNDER INCREASED INTRACRANIAL PRESSURE, MANNITOL (20%, 0.2G/ML), THE DOSE STATES 26.5 G (131.3 ML). WITH A 20% SOLUTION (0.2G/ML) AND DOSE OF 26.5G, THE DOSE IN VOLUME (ML) WOULD BE 132.5 ML. RESOURCE DOCUMENTS 131.3ML. BOTH VOLUMES ARE DOSAGES ARE WITHIN A DOSE RANGE FOR PATIENTS OF THESE WEIGHTS.