inforMED
MalfunctionIXY

GENESYS

Received Aug 27, 1999 · Event occurred Oct 22, 1998

Report 2916556-1999-00170 · MDR key 238113

Device

Generic name

Nuclear Medicine Equipment

Manufacturer

Adac Laboratories

Model number

2129-3011A

Catalog number

NA

Lot number

NA

Product problems

  • Mechanical Problem
  • Material Separation
  • Device Slipped
  • Other (for use when an appropriate device code cannot be identified)

Patient

*

Narrative

Description of Event or Problem

DETECTOR CAME OFF OF ITS MOUNT ON THE HEAD END STAND SIDE (HARMONIC DRIVE SIDE). WHEN THE RADIUS DRIVE SPROCKET SLID OFF OF ITS KEY, THE CHAIN FAILED TO DRIVE THE HEAD END OF THE DETECTOR WHILE THE FOOT END OF THE DETECTOR CONTINUED TO DRIVE. THIS CAUSED THE DETECTOR TO LEAN, PUTTING STRESS ON THE DETECTOR'S MOUNTING BRACKET.

Remedial action

  • Inspection