MalfunctionIXY
GENESYS
Received Aug 27, 1999 · Event occurred Oct 22, 1998
Report 2916556-1999-00170 · MDR key 238113
Device
Generic name
Nuclear Medicine Equipment
Manufacturer
Adac LaboratoriesModel number
2129-3011ACatalog number
NA
Lot number
NA
Product problems
- Mechanical Problem
- Material Separation
- Device Slipped
- Other (for use when an appropriate device code cannot be identified)
Patient
*
Narrative
Description of Event or Problem
DETECTOR CAME OFF OF ITS MOUNT ON THE HEAD END STAND SIDE (HARMONIC DRIVE SIDE). WHEN THE RADIUS DRIVE SPROCKET SLID OFF OF ITS KEY, THE CHAIN FAILED TO DRIVE THE HEAD END OF THE DETECTOR WHILE THE FOOT END OF THE DETECTOR CONTINUED TO DRIVE. THIS CAUSED THE DETECTOR TO LEAN, PUTTING STRESS ON THE DETECTOR'S MOUNTING BRACKET.
Remedial action
- Inspection