Received Dec 15, 2025 · Event occurred Oct 29, 2025

Report 1823260-2025-05122 · MDR key 23810641

Device

Generic name

Fluorometric, Cortisol

Manufacturer

Roche Diagnostics

Catalog number

07027150190

Lot number

885527

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Low Test Results
  • Non Reproducible Results
  • Incorrect, Inadequate or Imprecise Result or Readings
  • Low Test Results
  • Non Reproducible Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE ANALYZER SERIAL NUMBER IS (B)(6). THE INVESTIGATION IS ONGOING.

Description of Event or Problem

THERE WAS AN ALLEGATION OF QUESTIONABLE ELECSYS CORTISOL III RESULTS FOR 2 SAMPLES FROM THE SAME PATIENT ON A COBAS PURE E 402 ANALYTICAL UNIT. THE CUSTOMER QUESTIONED THE INITIAL LOW RESULTS WHICH PROMPTED THEM TO REPEAT THE PATIENT SAMPLES. SAMPLE 1 HAD AN INITIAL RESULT OF 26.2 NMOL/L. THE FIRST REPEAT RESULT WAS >1750 NMOL/L. A 1:10 DILUTION WAS PERFORMED ON THE SAMPLE AND THE RESULT WAS >17500 NMOL/L. A 1:100 DILUTION WAS PERFORMED ON THE SAMPLE AND THE RESULT WAS 20272 NMOL/L WITH FLAG. SAMPLE 2 HAD AN INITIAL RESULT OF 28.3 NMOL/L. THE REPEAT RESULT WAS 566 NMOL/L.

Additional Manufacturer Narrative

THE QC RECOVERY DATA PROVIDED WAS WITHIN SPECIFICATION. BASED ON THE QC DATA, A GENERAL REAGENT ISSUE COULD BE EXCLUDED. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED.