SMARTSET GHV GENTAMICIN 40G

Received Dec 12, 2025 · Event occurred Nov 25, 2025

Report 1818910-2025-21779 · MDR key 23793831

Device

Generic name

Bone Cement : Bone Cement

Catalog number

3095040

Lot number

4647943

Product problems

  • Chemical Problem
  • Chemical Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: BONE CEMENT WAS CLUMPED FROM THE BEGINNING, STILL IN THE BAG, CAN'T USE IN SURGICAL. THE SMARTSET GHV GENTAMICIN 40G (4647943) RETURNED TO DEPUY SYNTHES FOR EVALUATION. THE DEPUY SYNTHES TEAM CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. VISUAL ANALYSIS REVEALED THAT THE UNOPENED AMPULE AND THE CEMENT POUCH WERE RETURNED FOR EVALUATION. THE CEMENT POUCH WAS FOUND TO BE UNOPENED AND CONTAINED CEMENT WITH A HARDENED CONSISTENCY. NO OTHER DEFECTS THAT COULD HAVE CONTRIBUTED TO THE REPORTED EVENT WERE OBSERVED. A RETAINED SAMPLE TESTING WAS PERFORMED FOR THIS PRODUCT (3095040) AND LOT (4647943) COMBINATION. THE ASSESSMENT REVEALED THAT THE CEMENT MIXED AND BEHAVED AS EXPECTED FOR THE PRODUCT TYPE AND MET THE APPROPRIATE CONTROL SPECIFICATION. IT IS WORTH NOTICE THAT THE SAMPLE WAS TESTED IN A TEMPERATURE AND HUMIDITY-CONTROLLED LABORATORY. ADDITIONALLY, NO CLUMPING WAS OBSERVED IN THE SAMPLE POWDER PACK. BASED ON THE RESULTS OF THE RETAIN SAMPLE TESTING, THE ROOT CAUSE OF THE COMPLAINT CONDITION CANNOT BE TRACED TO A MANUFACTURING ISSUE. ONCE THE PRODUCT LEAVES DEPUY SYNTHES CONTROL, IT IS UNKNOWN WHAT ENVIRONMENT CONDITIONS OR HUMAN INTERVENTION OR MANIPULATION THE PRODUCTS ARE EXPOSED TO DURING THAT TIME. AS PER THE INSTRUCTIONS FOR USE (IFU-0630132 REV. B): "BONE CEMENTS ARE HEAT SENSITIVE. ANY INCREASE OR DECREASE IN TEMPERATURE (EITHER AMBIENT, AND/OR OF THE CEMENT COMPONENTS AND MIXING EQUIPMENT) FROM THE RECOMMENDED TEMPERATURE OF 73 °F (23 °C) WILL AFFECT THE HANDLING CHARACTERISTICS AND SETTING TIME OF THE CEMENT. NOTE: MANUAL HANDLING AND BODY TEMPERATURE WILL REDUCE THE FINAL SETTING TIME. VARIATIONS IN HUMIDITY WILL AFFECT THE CEMENT HANDLING CHARACTERISTICS AND SETTING TIME." ADDITIONALLY, " THE HANDLING CHARACTERISTICS AND SETTING TIME MAY VARY IF THE CEMENT COMPONENTS OR MIXING EQUIPMENT HAVE NOT BEEN FULLY EQUILIBRATED TO 73 °F (23 °C) BEFORE USE. IT IS RECOMMENDED THAT THE UNOPENED PRODUCT IS STORED AT 73 °F (23 °C) FOR A MINIMUM OF 24 HOURS BEFORE USE". HOWEVER, WITH THE AVAILABLE INFORMATION, A POTENTIAL CAUSE OF THE OBSERVED CONDITION OF THE CEMENT POWDER CANNOT BE ESTABLISHED. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT DESCRIPTION: SMARTSET GHV GENTAMICIN 40G PRODUCT CODE: 3095040 LOT NUMBER: 4647943 AND NO NON-CONFORMANCES OR MANUFACTURING IRREGULARITIES WERE IDENTIFIED THAT WOULD CONTRIBUTE TO THE REPORTED ISSUE. THE OVERALL COMPLAINT WAS CONFIRMED AS THE OBSERVED CONDITION OF THE SMARTSET GHV GENTAMICIN 40G WOULD CONTRIBUTE TO THE COMPLAINED DEVICE ISSUE. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF DEPUY SYNTHES QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE: 3095040 LOT NUMBER: 4647943 AND NO NON-CONFORMANCES OR MANUFACTURING IRREGULARITIES WERE IDENTIFIED THAT WOULD CONTRIBUTE TO THE REPORTED ISSUE. ADDED: D9. CORRECTED: H3.

Description of Event or Problem

IT WAS REPORTED THAT THE BONE CEMENT WAS CLUMPED FROM THE BEGINNING WHILE STILL IN THE BAG AND CAN'T BE USED IN A TOTAL KNEE ARTHROPLASTY (TKA) PROCEDURE. THERE WAS NO PATIENT CONSEQUENCE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. ADDITIONAL NARRATIVE: G4: DEVICE IS NOT DISTRIBUTED IN THE UNITED STATES BUT IS SIMILAR TO DEVICE MARKETED IN THE USA. DMF# - (B)(4) TRADE NAME ¿ GENTAMICIN SULPHATE ACTIVE INGREDIENT(S) ¿ GENTAMICIN SULPHATE DOSAGE FORM - POWDER STRENGTH ¿ 1.0G ACTIVE IN OUR CEMENTS. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative

PRODUCT COMPLAINT # (B)(4). THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES JOINT RECONSTRUCTION, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. H11 ADDITIONAL NARRATIVE: CORRECTED: D3, G1. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.