MalfunctionDXH
HOMECG+
Received Dec 12, 2025
Report 3012267350-2025-00018 · MDR key 23791389
Device
Generic name
Transmitters And Receivers, Electrocardiograph, Telephone
Manufacturer
Lohman Technologies, LlcModel number
V2.0Product problems
- Communication or Transmission Problem
- Communication or Transmission Problem
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Description of Event or Problem
THE PATIENT REPORTS THAT THE DEVICE BLINKS A BLANK SCREEN DESPITE CHANGING BATTERIES. THE PATIENT HAD NO ADVERSE EVENT OR REACTION RESULTING FROM THIS PROBLEM.
Additional Manufacturer Narrative
THE DEVICE IS UNABLE TO COMPLETE THE BOOT-UP PROCESS DUE TO A FAILURE OF THE INTERNAL FLASH ON THE MAIN PROCESSOR.
Remedial action
- Replace