TRUWAVE DISPOSABLE PRESSURE TRANSDUCER
Received Dec 12, 2025 · Event occurred Nov 17, 2025
Report 2015691-2025-09962 · MDR key 23787844
Device
Generic name
Catheter, Continuous Flush
Manufacturer
Edwards Lifesciences DrModel number
PX260Lot number
65997499
Product problems
- Delivered as Unsterile Product
- Delivered as Unsterile Product
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
CORRECTED SECTIONS D7A (SINGLE USE DEVICE REPROCESSED AND REUSED), E1 (REPORTER ADDRESS CITY AND PROVINCE), G2 (REPORT SOURCE, OTHER - CHINA) UPDATED SECTION H6 (COMPONENT CODE, TYPE OF INVESTIGATION, INVESTIGATION FINDINGS AND INVESTIGATION CONCLUSIONS). THE REPORT OF "POUCH NOT SEALED" WAS UNABLE TO BE CONFIRMED THROUGH THE PRODUCT EVALUATION, AS THE TYVEK POUCH WAS NOT RETURNED. AS RECEIVED, ALL CONNECTIONS WERE TIGHT AND SECURE. THE KIT WAS PRIMED AND FLUSHED WITHOUT ANY INDICATION OF FLOW RESTRICTION. NO LEAKAGE WAS DETECTED DURING LEAK TEST. THE DPT (DISPOSABLE PRESSURE TRANSDUCER) SENSOR ALSO ZEROED AND SENSED PRESSURE WITHOUT ANY PROBLEM. NO VISIBLE DAMAGE/DEFECT WAS OBSERVED FROM THE KIT DURING VISUAL EXAMINATION. THE MANUFACTURING RECORDS WERE REVIEWED FOR THE LOT INVOLVED AND THERE IS NO INDICATION OF A RELATED NONCONFORMANCE. ALL PROCESS PARAMETERS WERE MET AND INSPECTIONS PASSED SUCCESSFULLY. THERE ARE INTERNAL CONTROLS DURING THE MANUFACTURING PROCESS TO AVOID POTENTIAL CAUSES RELATED TO THE REPORTED MALFUNCTION, SUCH AS 100% VISUAL INSPECTION.
Description of Event or Problem
AS REPORTED, BEFORE USING THIS PRESSURE MONITORING SET, IT WAS FOUND THAT THE POUCH WAS NOT SEALED AND DID NOT MEET STERILE STANDARDS, THEREFORE, IT WAS REPLACED. NO PICTURE IS AVAILABLE. THERE IS NO ALLEGATION OF PATIENT INJURY. PATIENT DEMOGRAPHICS WERE NOT PROVIDED.
Additional Manufacturer Narrative
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS; HOWEVER, IT HAS NOT YET BEEN RECEIVED. UPON THE RETURN OF THE PRODUCT A SUPPLEMENTAL REPORT WILL BE SENT WITH THE INVESTIGATION RESULTS. COMPLAINT HISTORIES FOR ALL REPORTED EVENTS ARE REVIEWED AGAINST TRENDING CONTROL LIMITS ON A MONTHLY BASIS, AND ANY EXCURSIONS ABOVE THE CONTROL LIMITS ARE ASSESSED AND DOCUMENTED AS PART OF THIS MONTHLY REVIEW.