inforMED
MalfunctionOKI

BROSELOW RAINBOW TAPE

Received Dec 10, 2025 · Event occurred Dec 1, 2025

Report MW5180294 · MDR key 23771755

Device

Generic name

Emergency Response Safety Kit

Product problems

  • Application Program Problem: Dose Calculation Error

Patient

NA · Unknown

  • Insufficient Information
  • Insufficient Information

Narrative

Description of Event or Problem

AIRLIFE BRAND, REVISION 3, PRINT VERSION BROSELOW RAINBOW TAPE HAS TWO DOSING ERRORS, ONE SERIOUS. THIS IS THE TAPE THAT REPLACED THE RECALLED TAPE FROM MAY. VECURONIUM IV/IO (INTRAVENOUS (IV) /INTRAOSSEOUS (IO)) IS SHOWN AS 0.1 MG/ML INSTEAD OF 0.1 MG/KG. MORE CONCERNING IS THAT FLUMAZENIL IV/PO IS SHOWN AS 0.1 MG/KG INSTEAD OF 0.01 MG/KG. WE REPORTED THIS TO THE MANUFACTURER BUT WERE TOLD THERE ARE NO PLANS TO UPDATE AT THIS TIME.