inforMED
MalfunctionHST

HDS LONG OPERATIVE LEG SPAR

Received Dec 9, 2025 · Event occurred Nov 18, 2025

Report 3010216206-2025-00033 · MDR key 23763743

Device

Generic name

Apparatus, Traction, Non-powered

Model number

AR-6529-16

Catalog number

AR-6529-16

Lot number

N/A

Product problems

  • Unintended Movement
  • Unintended Movement

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

THE HIP DISTRACTOR SYSTEM IS DESIGNED TO POSITION, SUPPORT AND/OR DISTRACT THE PATIENT¿S HIP, POSTERIOR LOWER TORSO AND FOOT FOR HIP SURGERY. THESE DEVICES ARE INTENDED TO BE USED BY HEALTHCARE PROFESSIONALS WITHIN THE OPERATING ROOM SETTING. UPON INSPECTION, BAXTER FIELD SERVICE TECHNICIAN DETERMINED THAT THE STANDOFF HAD THREE SCREWS MISSING. A REPLACEMENT STANDOFF WITH SCREWS WERE REPLACED BY THE BAXTER TECHNICIAN. BASED ON THIS INFORMATION, NO FURTHER ACTIONS ARE REQUIRED AT THIS TIME. ALTHOUGH THERE WAS NO REPORTED INJURY WITH THIS EVENT, IF THE REPORT OF A DEVICE FALL DUE TO MISSING SCREWS WERE TO RECUR, IT COULD POTENTIALLY CAUSE SERIOUS INJURY OR DEATH. THEREFORE, BAXTER IS REPORTING THIS EVENT.

Description of Event or Problem

BAXTER RECEIVED A REPORT STATING THAT STABILIZER LEG HDS HAD STABILIZER LEG MISSING SCREWS AND CAUSED DEVICE TO FALL OUT. NO HARM OR SIGNIFICANT IMPACT TO THE SURGICAL PROCEDURE WAS REPORTED.

Additional Manufacturer Narrative

THE INVESTIGATION IS ONGOING. ALL ADDITIONAL AND RELEVANT INFORMATION THAT IS IDENTIFIED FOLLOWING COMPLETION OF THE INVESTIGATION WILL BE SUBMITTED IN A FINAL REPORT.