HYDROS ROBOTIC SYSTEM
Received Dec 9, 2025 · Event occurred Dec 4, 2025
Report 3012977056-2025-00296 · MDR key 23762238
Device
Generic name
Fluid Jet Removal System
Manufacturer
Procept Biorobotics CorporationCatalog number
HY1000T
Product problems
- Adverse Event Without Identified Device or Use Problem
- Adverse Event Without Identified Device or Use Problem
Patient
NA · Male
- Laceration(s)
- Laceration(s)
Narrative
Additional Manufacturer Narrative
THE HYDROS ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. THE INVESTIGATION CONSISTS OF INFORMATION RECEIVED, PLUS REVIEW OF THE TREATMENT LOGS, DEVICE HISTORY RECORD (DHR) AND LABELING. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) FOR HY1000/SERIAL NUMBER (B)(6) WAS CONDUCTED, WHICH CONFIRMED THAT THERE WERE NO NON-CONFORMANCES, FAILURES, DISCREPANCIES, OR MISSED STEPS DURING THE MANUFACTURING PROCESS THAT COULD BE RELATED TO THE REPORTED EVENT. THE REVIEW INDICATED THAT THE DEVICE MET ALL DESIGN AND MANUFACTURING SPECIFICATIONS WHEN RELEASED FOR DISTRIBUTION. THE LOG FILES WERE REVIEWED. NO INSTANCES OF ANY ERRORS WERE OBSERVED BEFORE, DURING, OR AFTER THE TREATMENT PASS, WHICH WAS COMPLETED SUCCESSFULLY. THE HYDROS ROBOTIC SYSTEM'S USER MANUAL (UM) UM0401-00-01, US, ENGLISH, REV. C, WAS REVIEWED. ALTHOUGH THE HYDROS ROBOTIC SYSTEM'S LABELING DOES NOT SPECIFICALLY MENTION DAMAGE TO URETERAL ORIFICES, PROCEPT'S RISK MANAGEMENT DOCUMENTATION INCLUDES DAMAGE TO URETERAL ORIFICES AS A CLINICAL EFFECT OF AQUABLATION THERAPY. THE HYDROS ROBOTIC SYSTEM IS A REUSABLE DEVICE; THEREFORE, IT IS STILL CURRENTLY IN POSSESSION OF THE USER FACILITY. IT WAS REPORTED THAT THE SURGEON HIT LEFT UO (URETERAL ORIFICE) WITH FIRST SWIPE OF LOOP DURING HEMOSTASIS. A STENT WAS PLACED. THE PATIENT HAS SINCE BEEN DISCHARGED AND DOING WELL. THE ROOT CAUSE OF THIS EVENT IS LIKELY USER ERROR. NO MALFUNCTION OF THE HYDROS ROBOTIC SYSTEM WAS REPORTED. ALTHOUGH THE HYDROS ROBOTIC SYSTEM'S LABELING DOES NOT SPECIFICALLY MENTION DAMAGE TO URETERAL ORIFICES, PROCEPT'S RISK MANAGEMENT DOCUMENTATION INCLUDES DAMAGE TO URETERAL ORIFICES AS A CLINICAL EFFECT OF AQUABLATION THERAPY. BASED ON THE REVIEW OF THE INFORMATION PROVIDED, PLUS A REVIEW OF THE TREATMENT LOG FILES, DHR AND UM, THE EVENT IS CONSIDERED NOT TO BE DEVICE RELATED. SUBMISSION OF THIS REPORT DOES NOT CONSTITUTE AN ADMISSION THAT THE MANUFACTURER'S PRODUCT CAUSED OR CONTRIBUTED TO THE EVENT.
Description of Event or Problem
A MALE PATIENT UNDERWENT AQUABLATION THERAPY FOR SYMPTOMATIC BENIGN PROSTATIC HYPERPLASIA (BPH). PROCEPT BIOROBOTICS (PROCEPT) BECAME AWARE THAT DURING HEMOSTASIS, THE SURGEON HIT THE LEFT URETERAL ORIFICE (UO) WITH THE FIRST SWIPE OF THE LOOP. THE TREATING SURGEON PLACED A STENT. THE PATIENT HAS SINCE BEEN DISCHARGED AND IS DOING WELL. NO MALFUNCTION OF THE HYDROS ROBOTIC SYSTEM WAS REPORTED.