inforMED
MalfunctionFMZ

GIRAFFE OMNIBED

Received Dec 9, 2025 · Event occurred Nov 11, 2025

Report 3036470679-2025-00017 · MDR key 23758371

Device

Generic name

Incubator, Neonatal

Manufacturer

Datex-ohmeda, Inc.

Catalog number

NA

Lot number

NA

Product problems

  • No Flow
  • No Flow

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

A GE HEALTHCARE SERVICE REPRESENTATIVE PERFORMED A CHECKOUT OF THE SYSTEM AND CONFIRMED THE REPORTED ISSUE. THE O2 SAFETY VALVE AND O2 FLOW VALVE WERE REPLACED TO RESOLVE THE ISSUE.

Additional Manufacturer Narrative

GE HEALTHCARE INVESTIGATION INTO THE REPORTED OCCURRENCE IS ONGOING. A FOLLOW-UP REPORT WILL BE ISSUED WHEN THE INVESTIGATION HAS BEEN COMPLETED. BLOCK A: NO REPORT OF PATIENT INVOLVEMENT. D4 PRIMARY UDI NUMBER: THERE IS NO UDI AVAILABLE AS THE DEVICE WAS MANUFACTURED PRIOR TO UDI COMPLIANCE REQUIREMENTS. LEGAL MANUFACTURER: GEHC TROUT - 3114 N GRANDVIEW BLVD. USA WAUKESHA, WI 53188.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A MALFUNCTION RESULTING IN LOSS OF AUXILIARY OXYGEN DURING SERVICE. THERE WAS NO PATIENT INVOLVEMENT.