ROCHE 9180 ELECTROLYTE ANALYZER
Received Dec 8, 2025 · Event occurred Nov 18, 2025
Report 1823260-2025-04997 · MDR key 23748327
Device
Product problems
- Incorrect Measurement
- Chemical Problem
- Device Ingredient or Reagent Problem
- Chemical Problem
- Device Ingredient or Reagent Problem
Patient
Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE ELECTRODE LOT NUMBERS AND EXPIRATION DATES WERE NOT PROVIDED. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE SODIUM AND LITHIUM RESULTS FROM THE ROCHE 9180 ELECTROLYTE ANALYZER. FOR SAMPLE "AAA" THE SODIUM RESULTS WERE 140 MMOL/L, 140 MMOL/L, 140 MMOL/L, 140 MMOL/L, 140 MMOL/L, 126 MMOL/L, 140 MMOL/L, AND 140 MMOL/L. THE LITHIUM RESULTS WERE 0.47 MMOL/L, 0.47 MMOL/L, 0.48 MMOL/L, 0.47 MMOL/L, 0.47 MMOL/L, 0.83 MMOL/L, 0.47 MMOL/L, AND 0.46 MMOL/L. FOR SAMPLE "BBB" THE SODIUM RESULTS WERE 126 MMOL/L, 140 MMOL/L, 140 MMOL/L, AND 140 MMOL/L. THE LITHIUM RESULTS WERE 0.68 MMOL/L, 0.32 MMOL/L, 0.33 MMOL/L, AND 0.33 MMOL/L.