ENVISTA SIMPLIFEYE INSERTER
Received Dec 8, 2025 · Event occurred Oct 9, 2025
Report 0001313525-2025-00221 · MDR key 23742803
Device
Generic name
Folders And Injectors, Intraocular Lens (iol)
Manufacturer
Bausch + LombModel number
21987Lot number
2304H10
Product problems
- Break
- Break
Patient
NA · Unknown
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
Narrative
Additional Manufacturer Narrative
THE HOSPITAL CONFIRMED THE PRODUCT IS NOT AVAILABLE FOR EVALUATION. A REVIEW OF DEVICE HISTORY RECORD (DHR) DID NOT IDENTIFY ANY ANOMALIES OR NON-CONFORMITIES THAT COULD BE RELATED TO THIS EVENT. THE LOT HISTORY, TREND ANALYSIS, RISK ANALYSIS AND/OR DIRECTIONS FOR USE REVIEW WERE CONSIDERED ACCEPTABLE, WITH THE PRODUCT PERFORMING WITHIN ANTICIPATED RATES. BASED ON THE AVAILABLE INFORMATION, THE ROOT CAUSE IS "DESIGN".
Description of Event or Problem
IT WAS REPORTED THAT DURING IMPLANTATION AN ADVERSE EVENT OCCURRED DURING SURGERY WHERE THE TIP OF AN ENVISTA INJECTOR SNAPPED INSIDE THE PATIENT¿S EYE. THE PATIENT REQUIRED AN ENLARGED INCISION (ABOUT 3.5 MM) AND REMOVAL OF THE FOREIGN BODY, FOLLOWED BY TWO STITCHES. THE ORIGINAL LENS REMAINED IN PLACE PRIOR TO THE INJECTOR TIP SNAPPING OFF, SO THAT LENS STAYED IN THE EYE. THE SURGEON'S OPINION ATTRIBUTES THE ISSUE TO THE ENVISTA INJECTOR. THE PATIENT HAS RECOVERED.