SODIUM ELECTRODE
Received Dec 5, 2025 · Event occurred Nov 13, 2025
Report 1823260-2025-04966 · MDR key 23729128
Device
Generic name
Electrode, Ion Specific, Sodium
Manufacturer
Roche DiagnosticsCatalog number
10825468001
Lot number
ASKU
Product problems
- Low Test Results
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
NO FURTHER ISSUES WERE REPORTED AFTER THE SERVICE ACTIONS WERE PERFORMED. THE INVESTIGATION COULD NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT COULD NOT BE DETERMINED.
Additional Manufacturer Narrative
THE COBAS 8000 ISE MODULE ANALYZER SERIAL NUMBER IS (B)(6). THE CUSTOMER STATED THERE MAY HAVE BEEN A DROP ON THE END OF A PROBE AND HE HAD NOT SEEN THIS BEFORE. THE FIELD SERVICE ENGINEER DETERMINED THERE WAS POSSIBLE CONTAMINATION OF THE ISE REAGENT FLOW PATH. ISE CHECKS REVEALED UNSTABLE VALUES. THE SYSTEM WAS DECONTAMINATED AND THE REAGENTS WERE PRIMED. THE VACUUM NOZZLE AND SOLENOIDS WERE CHECKED. ISE CHECKS WERE MORE STABLE AFTER THESE ACTIONS. CALIBRATION AND CONTROLS PASSED. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THE INITIAL REPORTER STATED THEY HAVE BEEN HAVING ONGOING ISSUES WITH CALIBRATION FAILURES AND QUALITY CONTROL RECOVERY ON THE COBAS 8000 ISE MODULE. A PHYSICIAN QUESTIONED SODIUM ELECTRODE RESULTS FOR AN UNSPECIFIED NUMBER OF PATIENT SAMPLES. THE CUSTOMER PROVIDED AN EXAMPLE OF ONE PATIENT SAMPLE WITH DISCREPANT SODIUM RESULTS. THE SAMPLE INITIALLY RESULTED IN A SODIUM VALUE OF 122 MMOL/L. AFTER THE PHYSICIAN QUESTIONED THE RESULT, THE SAMPLE WAS REPEATED ON AN UNKNOWN ANALYZER, RESULTING IN A VALUE OF 134 MMOL/L. THE REPEAT VALUE WAS DEEMED CORRECT.