inforMED
MalfunctionDQK

CARDIO 2

Received Aug 24, 1999 · Event occurred Aug 20, 1999

Report MW1017024 · MDR key 237117

Device

Generic name

Cardiopulmonary Stress Test Computer & Manual

Model number

UNK

Catalog number

UNK

Lot number

UNK

Product problems

  • Inadequate Instructions for Healthcare Professional
  • Application Program Problem: Parameter Calculation Error
  • Incorrect Or Inadequate Test Results

Patient

NA

Narrative

Description of Event or Problem

PHYSICIAN RUNNING TEST REQUESTED THAT MANUAL DATA CALCULATIONS BE DONE TO COMPARE AGAINST THE MACHINE'S SOFTWARE. THE RESPIRATORY THERAPIST REFERRED TO THE INSTRUCTION MANUAL FOR THE DEVICE AND FOUND THAT THE MFR HAD LEFT OUT CRITICAL PARENTHESES IN MOST OF THEIR PRINTED FORMULAS. IF USING THIS SET OF FORMULAS, THE USER WOULD MAKE EXTREME ERRORS IN CALCULATIONS WHICH MAY ADVERSELY AFFECT PT CARE. THE RPTR WAS ABLE TO OBTAIN ALTERNATIVE REFERENCES FOR THESE CALCULATIONS, BUT POINTS OUT THAT OTHERS MAY NOT EVEN RECOGNIZE THE ERROR. IT IS UNK WHETHER THE MANUAL CALCULATIONS CHECKED OUT AGAINST THE SOFTWARE DATA. RPTR BELIEVES MFR SHOULD HAVE NOTIFIED ALL OF ITS USERS.