inforMED
MalfunctionBZM

SENSORMEDICS

Received Aug 19, 1999

Report MW1016971 · MDR key 237110

Device

Generic name

Pulmonary Function Analyzer

Model number

VMAX 229

Catalog number

*

Lot number

*

Product problems

  • Incorrect Measurement
  • Repair
  • Replace
  • Other (for use when an appropriate device code cannot be identified)
  • Low Readings

Patient

*

  • Unknown (for use when the patient's condition is not known)

Narrative

Description of Event or Problem

SENSORMEDICS VMAX 229 PULMONARY FUNCTION SYSTEM CONSISTENTLY UNDERESTIMATES THE LUNG VOLUME DETERMINATION IN PTS WITH COPD. SYSTEM USES NITROGEN WASHOUT METHOD TO MEASURE LUNG VOLUME. REPLACED MODULES AND RECEIVED SOFTWARE UPGRADE WITH NO CHANGE IN RESULTS.