inforMED
MalfunctionCHI

LIPASE COLORIMETRIC ASSAY

Received Dec 2, 2025 · Event occurred Nov 8, 2025

Report 1823260-2025-04913 · MDR key 23696493

Device

Generic name

Lipase Test System

Manufacturer

Roche Diagnostics

Model number

08057982190

Catalog number

08057982190

Lot number

ASKU

Product problems

  • High Test Results
  • High Test Results

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Additional Manufacturer Narrative

BASED ON THE CONSISTENT RESULTS ACROSS TWO DIFFERENT COBAS C INSTRUMENTS, THE CAUSE OF THE EVENT WAS NOT AN INSTRUMENT OR REAGENT ISSUE. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT WAS CONSISTENT WITH A PHYSIOLOGICAL OR A PREANALYTICAL ISSUE AT THE CUSTOMER SITE. PREANALYTICS ARE WITHIN THE CUSTOMER'S RESPONSIBILITY.

Description of Event or Problem

WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH LIPASE COLORIMETRIC ASSAY ON A COBAS C 503 ANALYTICAL UNIT, NOT CONSISTENT WITH THE PATIENT'S CONDITION. INITIAL RESULT: 615 U/L. THE PHYSICIAN QUESTIONED THE INITIAL RESULT AS IT WAS HIGH AND WAS NOT CONSISTENT WITH THE PATIENT'S CONDITION. THE PHYSICIAN REQUESTED A NEW SAMPLE TO BE DRAWN AND TESTED ON THE SAME COBAS C 503 ANALYTICAL UNIT, RESULTING IN A VALUE OF 86 U/L. THE ORIGINAL SAMPLE WAS THEN REPEATED ON A DIFFERENT COBAS ANALYTICAL UNIT AT THE CUSTOMER'S SITE, AND THE VALUE OF 615 U/L WAS CONFIRMED. THE SPECIFIC RESULT RECEIVED WAS NOT PROVIDED.

Additional Manufacturer Narrative

THE LIPASE REAGENT EXPIRATION DATE WAS NOT PROVIDED. THE COBAS C 503 ANALYTICAL UNIT SERIAL NUMBER WAS: (B)(6). THE INVESTIGATION IS ONGOING.