LIPASE COLORIMETRIC ASSAY
Received Dec 2, 2025 · Event occurred Nov 8, 2025
Report 1823260-2025-04913 · MDR key 23696493
Device
Generic name
Lipase Test System
Manufacturer
Roche DiagnosticsModel number
08057982190Catalog number
08057982190
Lot number
ASKU
Product problems
- High Test Results
- High Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
BASED ON THE CONSISTENT RESULTS ACROSS TWO DIFFERENT COBAS C INSTRUMENTS, THE CAUSE OF THE EVENT WAS NOT AN INSTRUMENT OR REAGENT ISSUE. THE INVESTIGATION DID NOT IDENTIFY A PRODUCT PROBLEM. THE CAUSE OF THE EVENT WAS CONSISTENT WITH A PHYSIOLOGICAL OR A PREANALYTICAL ISSUE AT THE CUSTOMER SITE. PREANALYTICS ARE WITHIN THE CUSTOMER'S RESPONSIBILITY.
Description of Event or Problem
WE RECEIVED AN ALLEGATION ABOUT DISCREPANT RESULTS FOR 1 PATIENT SAMPLE TESTED WITH LIPASE COLORIMETRIC ASSAY ON A COBAS C 503 ANALYTICAL UNIT, NOT CONSISTENT WITH THE PATIENT'S CONDITION. INITIAL RESULT: 615 U/L. THE PHYSICIAN QUESTIONED THE INITIAL RESULT AS IT WAS HIGH AND WAS NOT CONSISTENT WITH THE PATIENT'S CONDITION. THE PHYSICIAN REQUESTED A NEW SAMPLE TO BE DRAWN AND TESTED ON THE SAME COBAS C 503 ANALYTICAL UNIT, RESULTING IN A VALUE OF 86 U/L. THE ORIGINAL SAMPLE WAS THEN REPEATED ON A DIFFERENT COBAS ANALYTICAL UNIT AT THE CUSTOMER'S SITE, AND THE VALUE OF 615 U/L WAS CONFIRMED. THE SPECIFIC RESULT RECEIVED WAS NOT PROVIDED.
Additional Manufacturer Narrative
THE LIPASE REAGENT EXPIRATION DATE WAS NOT PROVIDED. THE COBAS C 503 ANALYTICAL UNIT SERIAL NUMBER WAS: (B)(6). THE INVESTIGATION IS ONGOING.