CLEARLINK BLOOD RECIPIENT SET
Received Nov 26, 2025 · Event occurred Nov 2, 2025
Report 1416980-2025-05861 · MDR key 23660283
Device
Generic name
Set, Blood Transfusion
Manufacturer
Baxter Healthcare CorporationModel number
FNC1168NCatalog number
FNC1168N
Lot number
NI
Product problems
- Crack
- Fluid/Blood Leak
- Crack
- Fluid/Blood Leak
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WAS ADDED TO H6. H11: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, A PHOTOGRAPH OF THE SAMPLE WAS PROVIDED FOR EVALUATION. THE RETURNED PHOTOGRAPH WAS REVIEWED, AND IT WAS NOTED THAT THERE WAS A LEAK AT THE LEVEL OF THE HEAT SEAL CHAMBER. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE REPORTED CONDITION COULD NOT BE DETERMINED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Additional Manufacturer Narrative
SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
Description of Event or Problem
IT WAS REPORTED THAT AN UNSPECIFIED QUALITY OF BLOOD/SOLUTION ANESTHESIA/PCA INFUSION SETS WERE CRACKED WHICH RESULTED IN A FLUID LEAK. THE LEAKS WERE OBSERVED FROM THE DRIP CHAMBER, AND THE ISSUE WAS DESCRIBED AS ¿CRACK/BREAK IN RIDGED SET COMPONENTS¿. IT WAS UNKNOWN IF A PATIENT WAS CONNECTED AT THE TIME OF THESE EVENTS. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.