TOTAL PROTEIN URINE/CSF GEN.3
Received Nov 26, 2025 · Event occurred Oct 22, 2025
Report 1823260-2025-04882 · MDR key 23659510
Device
Generic name
Turbidimetric, Total Protein
Manufacturer
Roche DiagnosticsCatalog number
03333825190
Lot number
86777201
Product problems
- Non Reproducible Results
- Non Reproducible Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
THE COBAS INTEGRA 400 PLUS SERIAL NUMBER WAS (B)(6). BASED ON THE PROVIDED DATA, A GENERAL REAGENT PROBLEM COULD BE EXCLUDED BECAUSE CALIBRATION AND QUALITY CONTROL WERE ACCEPTABLE. THE INVESTIGATION IS ONGOING.
Description of Event or Problem
THERE WAS AN ALLEGATION OF QUESTIONABLE TOTAL PROTEIN URINE/CSF GEN.3 RESULTS FROM THE COBAS INTEGRA 400 PLUS ANALYZER. PATIENT 1 INITIAL RESULT WAS 275.03 MG/L, AND THE REPEAT RESULT IN ANOTHER LABORATORY WAS 82.1 MG/L. PATIENT 2 INITIAL RESULT WAS 105.5 MG/L, AND THE REPEAT RESULT IN ANOTHER LABORATORY WAS 68.1 MG/L.
Additional Manufacturer Narrative
DUE TO THE LIMITED INFORMATION PROVIDED, THE INVESTIGATION WAS UNABLE TO DETERMINE A SPECIFIC ROOT CAUSE. THERE WAS NO PRODUCT PROBLEM IDENTIFIED.
Additional Manufacturer Narrative
ADDITIONAL INFORMATION WAS PROVIDED FOR THE DATES OF THE QUESTIONABLE RESULTS. PATIENT 1 OCCURRED ON (B)(6) 2025. PATIENT 2 OCCURRED ON (B)(6) 2025. MEDWATCH FIELD B3 DATE OF EVENT WAS UPDATED.