BROSELOW TAPE 2025 EDITION 36-23446 REV.3

Received Nov 25, 2025 · Event occurred Nov 20, 2025

Report MW5179413 · MDR key 23646738

Device

Generic name

Emergency Response Safety Kit

Model number

36-23446 REV 3

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings
  • Device Markings/Labelling Problem

Patient

NA · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

INFORMATION CONTAINED ON THE AIRLIFE BROSELOW PEDIATRIC EMERGENCY REFERENCE TAPE IS INCORRECT FOR 2025 EDITION, 36-23446, REV3. IN THE CALCULATION BASIS TABLE, THE VECURONIUM DOSE LISTED UNDER "PRE-RSI/RSA NMB" IS LISTED AS 0.1MG/ML, THIS SHOULD BE LISTED AS 0.1MG/KG. ADDITIONALLY, THE DOSE LISTED FOR FLUMAZENIL IV/IO IS 0.1MG/KG, WHICH SHOULD BE 0.01MG/KG. THIS INFORMATION HAS POTENTIAL TO RESULT IN 10X DOSE ERROR FOR PEDIATRIC PATIENT.