Received Nov 24, 2025 · Event occurred Oct 27, 2025

Report 0001027229-2025-00002 · MDR key 23636452

Device

Generic name

Recliner Chair

Product problems

  • Improper or Incorrect Procedure or Method
  • Improper or Incorrect Procedure or Method

Patient

NA · Male

  • Head Injury
  • Head Injury

Narrative

Additional Manufacturer Narrative

THIS CHAIR WAS MANUFACTURED IN 2011 AND IS NO LONGER BEING SOLD. THE RECLINER OPERATED AS DESIGNED. THE ISSUE RESULTED FROM USER ERROR. THE UNIT WAS NOT RETURNED, AND NO FURTHER INVESTIGATION IS NEEDED, THIS INVESTIGATION IS CLOSED. NO FURTHER ACTIONS.

Description of Event or Problem

ON OCTOBER 27, IT WAS REPORTED BY (B)(6) THAT A PATIENT FELL FROM THE CHAIR. THE NURSE INTENDED TO ENGAGE THE FRONT LEFT CASTER BRAKE BUT INSTEAD ACCIDENTALLY UNLATCHED THE ROTARY LATCH (WHICH UNLOCKS THE SIDE OF THE CHAIR). BECAUSE THE PATIENT HAD MINIMAL TRUNK CONTROL, THIS ALLOWED THE PATIENT TO FALL FROM THE CHAIR WHEN THEY LEANED AGAINST THE ARM. THE PATIENT WAS SUBSEQUENTLY TRANSPORTED TO THE HOSPITAL FOR EVALUATION.