inforMED
MalfunctionOKI

BROSELOW TAPE

Received Nov 24, 2025 · Event occurred Nov 20, 2025

Report MW5179345 · MDR key 23631510

Device

Generic name

Emergency Response Safety Kit

Lot number

36-23446 REV 3

Product problems

  • Incorrect, Inadequate or Imprecise Result or Readings

Patient

12 MO · Unknown

  • No Clinical Signs, Symptoms or Conditions
  • No Clinical Signs, Symptoms or Conditions

Narrative

Description of Event or Problem

ERRORS ON BROSELOW TAPE MODEL 36-23446 REV 3 #1. FLUMAZENIL DOSING ON ONE OF THE DOSING TABLES TOWARDS THE BEGINNING OF THE CARD LISTS THE DOSE AS 0.1 MG/KG. (THE DOSE SHOULD BE 0.01 MG/KG). #2. VECURONIUM ON ONE OF THE DOSING TABLES TOWARDS THE BEGINNING OF THE CARD, FOR THE "PRE-RSI" INDICATION, LISTS THE DOSE AS 0.1 MG/ML (SHOULD BE MG/KG). - THE VECURONIUM DOSE UNDER THE POST-RSI IS CORRECT RIGHT BELOW IT.