DXC 700 AU CHEMISTRY ANALYZER
Received Nov 21, 2025 · Event occurred Nov 3, 2025
Report 9612296-2025-01016 · MDR key 23624463
Device
Generic name
Analyzer, Chemistry (photometric, Discrete), For Clinical Use
Manufacturer
Beckman Coulter Mishima K.k.Catalog number
B86444
Product problems
- Low Test Results
- Low Test Results
Patient
NA · Unknown
- No Clinical Signs, Symptoms or Conditions
- No Clinical Signs, Symptoms or Conditions
Narrative
Additional Manufacturer Narrative
BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE DXC 700 AU CHEMISTRY ANALYZER AND IDENTIFIED THE PROBABLE CAUSE AS A LEAKING VALVE ASSEMBLY. THE FSE REPLACED THE SOLENOID VALVE TO RESOLVE THE ISSUE. SECTION A2, A3, A4, A5 AND A6: INFORMATION NOT PROVIDED BY CUSTOMER. THE BECKMAN COULTER INTERNAL IDENTIFIER IS(B)(4).
Description of Event or Problem
THE CUSTOMER REPORTED THE GENERATION OF ERRONEOUSLY LOW PATIENT RESULTS FOR RHEUMATOID FACTOR (RF), ALANINE AMINOTRANSFERASE (ALT), TOTAL BILIRUBIN (TBIL), CREATININE (CREA), UREA NITROGEN (BUN), AND POTASSIUM (K) ON THEIR DXC700AU CHEMISTRY ANALYZER. THERE WAS NO REPORT OF CHANGE TO TREATMENT, INJURY, OR DEATH ASSOCIATED WITH EVENT. THE CUSTOMER DID NOT PROVIDE PATIENT DATA OR DEMOGRAPHICS.