inforMED
MalfunctionIKZ

ISOLIBRIUM PE SUPPORT SURFACE

Received Nov 21, 2025 · Event occurred May 16, 2025

Report 0001831750-2025-00747 · MDR key 23620818

Add to watch list
Product code (IKZ)
Model (297300000000)

Device

Generic name

Bed, Patient Rotation, Powered

Catalog number

297300000000

Product problems

  • Positioning Problem
  • Material Too Soft/Flexible
  • Positioning Problem
  • Material Too Soft/Flexible

Patient

NA · Unknown

  • Pressure Sore/Ulcer
  • Pressure Sore/Ulcer

Narrative

Additional Manufacturer Narrative

FURTHER INVESTIGATION IDENTIFIED THAT THE REPORTED ISSUES WERE DUE TO PUMP HOSES COMING DETACHED FROM THE MATTRESS AS WELL AS THE STAFF NOT SETTING THE PATIENT WEIGHT CORRECTLY. BASED ON AVAILABLE INFORMATION, STRYKER DETERMINED THAT THE PRESSURE INJURY WAS NOT SERIOUS IN NATURE.

Description of Event or Problem

NO NEW INFORMATION.

Description of Event or Problem

IT WAS REPORTED THAT THERE WAS A PRESSURE INJURY DUE TO THE MATTRESS ADJUSTING THE PATIENT'S POSITIONING/WEIGHT DIFFERENTLY THAN THE CUSTOMER HAD SET IT TO.